Regular Article
Evaluation of Measures of Technical Image Quality for Intracranial Magnetic Resonance Angiography

https://doi.org/10.1006/cbmr.1999.1526Get rights and content

Abstract

We evaluate three measures of technical image quality for intracranial magnetic resonance angiography (MRA): (1) a two-alternative forced choice (2AFC) evaluation of vessel visibility, (2) vessel-to-background signal-difference-to-noise ratio (SDNR), and (3) observer ranking of the fidelity of vessel morphology compared to that in a gold standard image. The gold standard used for both the 2AFC and ranking measures is intraarterial catheter angiography. These measures are applied to healthy arterial segments. The 2AFC and SDNR measures directly evaluate the visibility of artery segments for which the existence is known from the gold standard images. We argue that (1) 2AFC evaluates the carrier signals on which any vascular disease process is modulated and provides an upper bound on the detectibility of vascular lesions, (2) SDNR is a predictor of 2AFC, and (3) ranking may be used to predict the relative performance of techniques in the detection of vascular lesions.

References (30)

  • H.H. Barrett et al.

    Linear discriminants and image quality

    Image Vision Comput.

    (1992)
  • C.E. Metz

    Basic principles of ROC analysis

    Sem. Nucl. Med.

    (1978)
  • H.H. Barret et al.

    Model observers for assessment of image quality

    Proc. Natl. Acad. Sci. USA

    (1993)
  • J.A. Swets

    ROC analysis applied to the evaluation of medical imaging techniques

    Invest. Radiol.

    (1979)
  • J.A. Hanley et al.

    The meaning and use of the area under a receiver operating characteristic (ROC) curve

    Radiology

    (1982)
  • J.A. Swets

    Measuring the accuracy of diagnostic systems

    Science

    (1988)
  • A.E. Burgess

    Comparison of receiver operating characteristic and forced choice observer performance measurement methods

    Med. Phys.

    (1995)
  • M.P. Eckstein et al.

    Visual signal detection in structured backgrounds I. Effect of number of possible spatial locations and signal contrast

    J. Opt. Soc. Am. A

    (1996)
  • D.G. Fryback et al.

    The efficacy of diagnostic imaging

    Med. Decis. Making

    (1991)
  • H.R. Blackwell

    Contrast thresholds of the human eye

    J. Opt. Soc. Am.

    (1946)
  • H.B. Barlow

    Increment thresholds at low intensities

    J. Physiol.

    (1957)
  • A. Rose

    Vision: Human and Electronic

    (1973)
  • Y.P. Du et al.

    Contrast-to-noise ratio measurements in 3D MR angiography

    Invest. Radiol.

    (1993)
  • Y.P. Du et al.

    Experimental and theoretical studies of vessel contrast-to-noise ratio in intracranial time-of-flight MR angiography

    J. Magn. Reson. Imaging

    (1996)
  • This work was supported by NIH Grant ROI HL53596.

    View full text