Abstract
Physics and radiographer QC procedures are urgently needed as the first few contrast-enhanced dual energy systems have been installed in the U.K. Preliminary work on one commercially available system has enabled us to propose new tests, relevant to the properties of dual energy imaging systems. Results are presented for measurements with the chosen phantom, which contains disks with a range of iodine content from 0.25 to 2 mg/cm2. Breasts of different thicknesses and different glandularity were simulated by adding slabs of CIRS material, of a range of compositions, on top of the phantom. The system tested had a response which was proportional to the iodine content of disks in the phantom, had good reproducibility, and did not change significantly when simulated breast thickness and composition were varied.
Keywords
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
References
Kulama, E., Burch, A., Castellano, I., Lawinski, C.P., Marshall, N., Young, K.C.: Commissioning and Routine Testing of Full Field Digital Mammography Systems (NHSBSP Equipment Report 0604). NHS Cancer Screening Programmes, Sheffield (2009)
Van Engen, R., Young, K.C., Bosmans, H., Thijssen, M.: The European Protocol for the Quality Control of the Physical and Technical Aspects of Mammography Screening. In: European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edn. European Commission, Luxembourg (2006)
Leithner, R., Knogler, T., Homolka, P.: Development and Production of a Prototype Iodine Contrast Phantom for CEDEM. Phys. Med. Biol. 58, N25–N35 (2013)
Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., Kotre, C.J.: Additional Factors for the Estimation of Mean Glandular Breast Dose using the UK Mammography Dosimetry Protocol. Phys. Med. Biol. 45, 3225–3240 (2000)
Dance, D.R., Young, K.C.: Estimation of Mean Glandular Dose for Contrast Enhanced Digital Mammography: Factors for use with the UK, European and IAEA Breast Dosimetry Protocols. Phys. Med. Biol. 59(9), 2127–2137
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2014 Springer International Publishing Switzerland
About this paper
Cite this paper
Oduko, J., Homolka, P., Jones, V., Whitwam, D. (2014). A Protocol for Quality Control Testing for Contrast-Enhanced Dual Energy Mammography Systems. In: Fujita, H., Hara, T., Muramatsu, C. (eds) Breast Imaging. IWDM 2014. Lecture Notes in Computer Science, vol 8539. Springer, Cham. https://doi.org/10.1007/978-3-319-07887-8_57
Download citation
DOI: https://doi.org/10.1007/978-3-319-07887-8_57
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-07886-1
Online ISBN: 978-3-319-07887-8
eBook Packages: Computer ScienceComputer Science (R0)