Abstract
We propose a framework using freely available tools for the synthesis of physically realistic CT measurements for low dose reconstruction development and validation, using a fully sampled Monte Carlo method. This allows the generation of test data that has artefacts such as photon starvation, beam-hardening and scatter, that are both physically realistic and not unfairly biased towards model-based iterative reconstruction (MBIR) algorithms. Using the open source Monte Carlo tool GATE and spectrum simulator SpekCalc, we describe how physical elements such as source, specimen and detector may be modelled, and demonstrate the construction of fan-beam and cone-beam CT systems. We then show how this data may be consolidated and used with image reconstruction tools. We give examples with a low dose polyenergetic source, and quantitatively analyse reconstructions against the numerical ground-truth for MBIR with simulated and ‘inverse crime’ data. The proposed framework offers a flexible and easily reproducible tool to aid MBIR development, and may reduce the gap between synthetic and clinical results.
Keywords
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
Liu, L.: Reconstruction, model-based iterative: a promising algorithm for today’s computed tomography imaging. J. Med. Imaging Radiat. Sci. 45(2), 131–136 (2014)
Sidky, E.Y., Kao, C.M., Pan, X.: Accurate image reconstruction from few-views and limited-angle data in divergent-beam CT. J. X-ray. Sci. Technol. 14(2), 1–30 (2006)
Hu, Z., Liang, D., Xia, D., Zheng, H.: Compressive sampling in computed tomography: method and application. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip. 748, 26–32 (2014)
Candes, E.J., Romberg, J., Tao, T.: Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information. IEEE Trans. Inf. Theory 52(2), 489–509 (2006)
Donoho, D.L.: Compressed sensing. IEEE Trans. Inf. Theory 52, 1289–1306 (2006)
Pan, X., Sidky, E.Y., Vannier, M.: Why do commercial CT scanners still employ traditional, filtered back-projection for image reconstruction? Inverse Probl. 25, 123009 (2009)
Kaipio, J., Somersalo, E.: Statistical inverse problems: discretization, model reduction and inverse crimes. J. Comput. Appl. Math. 198(2), 493–504 (2007)
Nuyts, J., De Man, B., Fessler, J.A., Zbijewski, W., Beekman, F.J.: Modelling the physics in the iterative reconstruction for transmission computed tomography. Phys. Med. Biol. 58(12), R63–R96 (2013)
De Man, B., Basu, S., Chandra, N., Dunham, B., Edic, P., Iatrou, M., McOlash, S., Sainath, P., Shaughnessy, C., Tower, B., Williams, E.: CatSim: a new computer assisted tomography simulation environment, p. 65102G, March 2007
Jia, X., Yan, H., Cerviño, L., Folkerts, M., Jiang, S.B.: A GPU tool for efficient, accurate, and realistic simulation of cone beam CT projections. Med. Phys. 39(12), 7368–7378 (2012)
Jan, S., Benoit, D., Becheva, E., Carlier, T., Cassol, F., Descourt, P., Frisson, T., Grevillot, L., Guigues, L., Maigne, L., Morel, C., Perrot, Y., Rehfeld, N., Sarrut, D., Schaart, D.R., Stute, S., Pietrzyk, U., Visvikis, D., Zahra, N., Buvat, I.: GATE V6: a major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy. Phys. Med. Biol. 56(4), 881–901 (2011)
Poludniowski, G., Landry, G., DeBlois, F., Evans, P.M., Verhaegen, F.: SpekCalc: a program to calculate photon spectra from tungsten anode X-ray tubes. Phys. Med. Biol. 54(19), N433–N438 (2009)
ICRP Publication 110: Adult reference computational phantoms. Ann. ICRP 39(2), 1 (2009)
Fessler, J.A.: Image Reconstruction Toobox, 14 November 2014
Saloman, E.B., Hubbell, J.H., Scofield, J.H.: X-ray attenuation cross sections for energies 100 eV to 100 keV and elements Z = 1 to Z = 92. At. Data Nucl. Data Tables 38(1), 1–196 (1988)
Chang, Z., Zhang, R., Thibault, J.-B., Sauer, K., Bouman, C.: Statistical X-ray computed tomography imaging from photon-starved measurements. SPIE Comput. Imaging 9020, 90200G (2014)
Elbakri, I.A., Fessler, J.A.: Statistical image reconstruction for polyenergetic X-ray computed tomography. IEEE Trans. Med. Imaging 21(2), 89–99 (2002)
Barrett, J.F., Keat, N.: Artifacts in CT: recognition and avoidance. RadioGraphics 24(6), 1679–1691 (2004)
Sun, M., Star-Lack, J.M.: Improved scatter correction using adaptive scatter kernel superposition. Phys. Med. Biol. 55(22), 6695–6720 (2010)
Perraudin, N., Kalofolias, V., Shuman, D., Vandergheynst, P.: UNLocBoX: A MATLAB convex optimization toolbox for proximal-splitting methods, February 2014
Schneider, U., Pedroni, E., Lomax, A.: The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys. Med. Biol. 41(1), 111–124 (1996)
Siewerdsen, J.H., Jaffray, D.A.: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med. Phys. 28(2), 220 (2001)
Acknowledgements
This work was supported by the Maxwell Advanced Technology Fund, EPSRC DTP studentship funds and ERC project: C-SENSE (ERC-ADG-2015-694888). MD is also supported by a Royal Society Wolfson Research Merit Award.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2017 Springer International Publishing AG
About this paper
Cite this paper
Mason, J.H., Nailon, W.H., Davies, M.E. (2017). A Monte Carlo Framework for Low Dose CT Reconstruction Testing. In: Tsaftaris, S., Gooya, A., Frangi, A., Prince, J. (eds) Simulation and Synthesis in Medical Imaging. SASHIMI 2017. Lecture Notes in Computer Science(), vol 10557. Springer, Cham. https://doi.org/10.1007/978-3-319-68127-6_9