Skip to main content

Image Quality Assessment and Equipment Optimisation with Automated Phantom Evaluation in Full Field Digital Mammography (FFDM)

  • Conference paper

Part of the book series: Lecture Notes in Computer Science ((LNIP,volume 5116))

Abstract

The work investigates the feasibility of using a commercial software performing automatic threshold-detail-image analysis for absolute image quality evaluation (phantom CDMAM ver. 4.0 + CDMAM Analyser ver. 1.1).

The software proposes an overall image quality index IQFinv. The quantities IQFinv, contrast-to-noise-ratio (CNR), threshold contrast and average glandular dose (AGD) are studied for a wide range of exposure settings in three full field digital mammographic units of different manufacturers. IQFinv correlates linearly with CNR and exhibits a better correlation to threshold contrast than CNR itself in all units, turning out to be a reliable parameter for image quality assessment.

At comparable AGD and exposure settings, all systems exhibit equivalent image quality. Using the automatic exposure control (AEC) all units achieve, for thin and medium breasts, a higher image quality than requested by the European Guidelines, whereas only in one unit the image quality remains above the desired level for thick breasts.

This is a preview of subscription content, log in via an institution.

Buying options

Chapter
USD   29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD   84.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD   109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Learn about institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Burgess, A.E.: The Rose Model, revisted. J. Optical Soc. Am. 16, 633–646 (1999)

    Article  Google Scholar 

  2. Van Engen, R., Young, K., Bonmans, H., Thijssen, M.: European Protocol for the Quality Control of the Physical and Technical Aspects of Mammography Screening. Part B: Digital Mammography. In: European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edn., European Communities, Luxembourg (2006)

    Google Scholar 

  3. Jaffe, M.J., Jong, R.A.: Image Quality in Digital Mammography. In: Samei, E., Durham, N.C. (eds.) Syllabus. Advances in Digital Radiography. RSNA Categorical Course in Diagnostic Radiology Physics 2003, RSNA, Chicago, vol. 2003, pp. 123–141 (2003)

    Google Scholar 

  4. IEC 62220-1-2: Medical Electrical Equipment - Characteristics of Digital X-ray Imaging Devices. Part 1-2: Determination of the Detective Quantum Efficiency – Detectors used in mammography, IEC,Geneva (2007)

    Google Scholar 

  5. Aufrichtig, R., Xue, P.: Dose Efficiency and Low-Contrast detectability of an Amorphous Silicon X-ray Detecor for Digital Radiography. Phys. Med. Biol. 45, 2653–2669 (2000)

    Article  Google Scholar 

  6. Workman, A., Castellano, I., Kulama, E., Lawinski, C.P., Marshall, N., Young, K.C.: Commissioning and Routine Testing of Full Field Digital Mammography Systems. NHSBSP Equipment Report 0604, ver. 2. NHS Cancer Screening Programmes (2006), http://www.cancerscreening.nhs.uk

  7. Young, K.C., Cook, J., Odoku, J.M.: Use of the European Protocol to Optimise a Digital Mammography System. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds.) IWDM 2006. LNCS, vol. 4046, pp. 362–369. Springer, Heidelberg (2006)

    Chapter  Google Scholar 

  8. Young, K.C., Cook, J., Odoku, J.M., Bosnans, H.: Comparison of Software and Human Observers in Reading Images of the CDMAM Test Object to Assess Digital Mammography Systems. In: Flynn, M.J., Hsieh, J. (eds.) Proc. SPIE Medical Imaging 2006: Physics of Medical Imaging, vol. 6142, p. 614206 (2006)

    Google Scholar 

  9. Young, K.C., Oduko, J.M., Bosmans, H., Nijs, K., Martinez, L.: Optimal beam quality selection in digital mammography. Br. J. Radiol. 79, 981–990 (2006)

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Elizabeth A. Krupinski

Rights and permissions

Reprints and permissions

Copyright information

© 2008 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Oberhofer, N., Paruccini, N., Moroder, E. (2008). Image Quality Assessment and Equipment Optimisation with Automated Phantom Evaluation in Full Field Digital Mammography (FFDM). In: Krupinski, E.A. (eds) Digital Mammography. IWDM 2008. Lecture Notes in Computer Science, vol 5116. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-70538-3_33

Download citation

  • DOI: https://doi.org/10.1007/978-3-540-70538-3_33

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-70537-6

  • Online ISBN: 978-3-540-70538-3

  • eBook Packages: Computer ScienceComputer Science (R0)

Publish with us

Policies and ethics