Estimation of the effects of inset heights and slit configurations in an acetabular cup on the pull-out behavior of an artificial hip joint with a structure for preventing dislocation using finite element analysis

  • Original Article
  • Published:
Medical & Biological Engineering & Computing Aims and scope Submit manuscript

Abstract

Joint dislocation is a critical problem of total hip replacement. We have newly proposed an artificial hip joint with a structure that prevents dislocation. The proposed joint has a simple form with a femoral head partially covered with an acetabular cup. In the present study, the effects of inset heights and slit configurations of the cup on the pull-out forces of the joint were evaluated using finite element analysis. Joint models with different inset heights and those with or without a slit in the cup were used for the analyses to estimate the pull-out forces of the joint. In the case without the slit, the maximum pull-out force of the joint with 1.0 and 1.5 mm of the inset height was approximately 12 and 40 N, respectively. In the case of 1.0-mm inset height, the maximum force of the joint with and without the slit was approximately 9 and 12 N, respectively. These results reveal that the maximum force is markedly changed by the inset height and is moderately affected by the slit. Thus, we can gain insights into a strategy to optimally design an artificial joint in which dislocation does not occur easily.

Graphical abstract

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12

Similar content being viewed by others

References

  1. Colic K, Sedmak A (2016) The current approach to research and design of the artificial hip prosthesis: a review. Rheumatol Orthop Med 1:1–7. https://doi.org/10.15761/ROM.1000106

    Article  Google Scholar 

  2. Affatato S, Ruggiero A, Merola M (2015) Advanced biomaterials in hip joint arthroplasty. A review on polymer and ceramics composites as alternative bearings. Compos B Eng 83:276–283. https://doi.org/10.1016/j.compositesb.2015.07.019

    Article  CAS  Google Scholar 

  3. Ghosh S, Abanteriba S (2016) Status of surface modification technique for artificial hip implants. Sci Technol Adv Mater 17:715–735. https://doi.org/10.1080/14686996.2016.1240575

    Article  PubMed  PubMed Central  Google Scholar 

  4. Seagrave KG, Troelsen A, Malchau H, Husted H, Gromov K (2017) Acetabular cup position and risk of dislocation in primary total hip arthroplasty: a systematic review of the literature. Acta Orthop 88:10–17. https://doi.org/10.1080/17453674.2016.1251255

    Article  PubMed  Google Scholar 

  5. Summer DR (2015) Long-term implant fixation and stress-shielding in total hip replacement. J Biomech 48:797–800. https://doi.org/10.1016/j.jbiomech.2014.12.021

    Article  Google Scholar 

  6. Charnley J (1970) Total hip replacement by low-friction arthroplasty. Clin Orthop Relat Res 72:7–21

    CAS  PubMed  Google Scholar 

  7. Masaoka T, Yamamoto K, Shishido T, Katori Y, Mizoue T, Shirasu H, Nunoda D (2006) Study of hip joint dislocation after total hip arthroplasty. Int Orthop 30:26–30. https://doi.org/10.1007/s00264-005-0032-4

    Article  PubMed  Google Scholar 

  8. Wetters NG, Murray TG, Moric M, Sporer SM, Paprosky WG, Della Valle CJ (2013) Risk factors for dislocation after revision total hip arthroplasty. Clin Orthop Relat Res 471:410–416. https://doi.org/10.1007/s11999-012-2561-7

    Article  PubMed  Google Scholar 

  9. Marcellin-Little DJ, DeYoung BA, Doyens DH, DeYoung DJ (1999) Canine uncemented porous-coated anatomic total hip arthroplasty: results of a long-term prospective evaluation of 50 consecutive cases. Vet Surg 28:10–20. https://doi.org/10.1053/jvet.1999.0010

    Article  CAS  PubMed  Google Scholar 

  10. Montgomery RD, Milton JL, Pernell R, Aberman HM (1992) Total hip arthroplasty for treatment of canine hip dysplasia. Vet Clin North Am Small Anim Pract 22:703–719. https://doi.org/10.1016/S0195-5616(92)50063-8

    Article  CAS  PubMed  Google Scholar 

  11. Liska WD, Doyle N, Marcellin-Little DJ, Osborne A (2009) Total hip replacement in three cats: surgical technique, short-term outcome and comparison to femoral head osteotomy. Vet Comp Orthop Traumatol 22:505–510. https://doi.org/10.3415/VCOT-08-09-0087

    Article  CAS  PubMed  Google Scholar 

  12. Huggons N, Andrea R, Grant C, Duncan C (2010) Total hip arthroplasty in the horse: overview, technical considerations and case report. Equine Vet Educ 22:547–553. https://doi.org/10.1111/j.2042-3292.2010.00118.x

    Article  Google Scholar 

  13. Dyce J, Wisner ER, Wang Q, Olmstead ML (2000) Evaluation of risk factors for luxation after total hip replacement in dogs. Vet Surg 29:524–532. https://doi.org/10.1053/jvet.2000.17858

    Article  CAS  PubMed  Google Scholar 

  14. Brown DC, Boston RC, Farrar JT (2013) Comparison of force plate gait analysis and owner assessment of pain using the canine brief pain inventory in dogs with osteoarthritis. J Vet Intern Med 27:22–30. https://doi.org/10.1111/jvim.12004

    Article  CAS  PubMed  Google Scholar 

  15. Carr BJ, Dycus DL (2016) Canine gait analysis. Today Vet Pract 6:93–100

    Google Scholar 

  16. Bartz RL, Nobel PC, Kadakia NR, Tullos HS (2000) The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am 82:1300–1307. https://doi.org/10.2106/00004623-200009000-00010

    Article  CAS  PubMed  Google Scholar 

  17. Chi WM, Lin CC, Ho YJ, Lin HC, Chen JM (2018) Using nonlinear finite element models to analyse stress distribution and torque required for dislocation of newly developed total hip structure after prosthetic impingement. Med Biol Eng Comput 56:37–47. https://doi.org/10.1007/s11517-017-1673-3

    Article  PubMed  Google Scholar 

  18. Shapiro GS, Weiland DE, Markel DC, Padgett DE, Sculco TP, Pellicci PM (2003) The use of a constrained acetabular component for recurrent dislocation. J Arthroplast 18:250–258. https://doi.org/10.1054/arth.2003.50090

    Article  Google Scholar 

  19. Martino ID, Triantafyllopoulos GK, Sculco PK, Sculco TP (2014) Dual mobility cups in total hip arthroplasty. World J Orthop 5:180–187. https://doi.org/10.5312/wjo.v5.i3.180

    Article  PubMed  PubMed Central  Google Scholar 

  20. Belytschko T, Gracie R, Ventura G (2009) A review of extended/generalized finite element methods for material modeling. Model Simul Mater Sci Eng 17:1–24. https://doi.org/10.1088/0965-0393/17/4/043001

    Article  Google Scholar 

  21. Charmpis DC, Schueller GI, Pellissetti MF (2007) The need for linking micromechanics of materials with stochastic finite elements: a challenge for materials science. Comput Mater Sci 41:27–37. https://doi.org/10.1016/j.commatsci.2007.02.014

    Article  Google Scholar 

  22. Fries TP, Belytschko T (2010) The extended/generalized finite element method: an overview of the method and its applications. Int J Numer Methods Eng 84:253–304. https://doi.org/10.1002/nme.2914

    Article  Google Scholar 

  23. Huebner LM, Thornton MB, Byrom AK (1995) The finite element method for engineers. Wiley, New York

    Google Scholar 

  24. Geng JP, Tan KB, Liu GR (2001) Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 85:585–598. https://doi.org/10.1067/mpr.2001.115251

    Article  CAS  PubMed  Google Scholar 

  25. Hirata Y, Inaba Y, Kobayashi N, Ike H, Fujimaki H, Saito T (2013) Comparison of mechanical stress and change in bone mineral density between two types of femoral implant using finite element analysis. J Arthroplast 28:1731–1735. https://doi.org/10.1016/j.arth.2013.04.034

    Article  Google Scholar 

  26. Morioka D, Kitayama I, Nakano K, Kitano M, Yamanaka T, Koyama H, Morimoto T, Sonobe H, Miyazaki N (2016) Finite element analysis of trimmed plastic ankle-foot orthoses. Adv Exp Mech 1:269–276. https://doi.org/10.11395/aem.1.0_269

    Article  Google Scholar 

  27. Qian L, Todo M, Matsushita Y, Koyano K (2009) Finite element analysis of bone resorption around dental implant. J Biomech Sci Eng 4:365–376. https://doi.org/10.1299/jbse.4.365

    Article  Google Scholar 

  28. Rebelo N, Perry M (2000) Finite element analysis for the design of nitinol medical devices. Minim Invasive Ther Allied Technol 9:75–80. https://doi.org/10.3109/13645700009063053

    Article  Google Scholar 

  29. Valero C, Javierre E, Garcia-Aznar JM, Gomez-Benito MJ (2014) Nonlinear finite element simulations of injuries with free boundaries: application to surgical wounds. Int J Numer Method Biomed Eng 30:616–633. https://doi.org/10.1002/cnm.2621

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Al-Dirini RMA, Martelli S, Huff D, Zhang J, Clement JG, Besier T, Taylor M (2018) Evaluating the primary stability of standard vs lateralised cementless femoral stems-a finite element study using a diverse patient cohort. Clin Biomech 59:101–109. https://doi.org/10.1016/j.clinbiomech.2018.09.002

    Article  Google Scholar 

  31. Ashkanfar A, Langton DJ, Joyce TJ (2017) A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: a finite element wear analysis. J Mech Behav Biomed Mater 69:257–266. https://doi.org/10.1016/j.jmbbm.2017.01.018

    Article  PubMed  Google Scholar 

  32. Dopico-Gonzalez C, New AM, Browne M (2010) Probabilistic finite element analysis of the uncemented hip replacement-effect of femur characteristics and implant design geometry. J Biomech 43:512–520. https://doi.org/10.1016/j.jbiomech.2009.09.039

    Article  PubMed  Google Scholar 

  33. Ezquerra L, Quilez MP, Perez MA, Albareda J, Seral B (2017) Range of movement for impingement and dislocation avoidance in total hip replacement predicted by finite element model. J Med Biol Eng 37:26–34. https://doi.org/10.1007/s40846-016-0210-4

    Article  PubMed  PubMed Central  Google Scholar 

  34. Tonino A, Oosterbos C, Rahmy A, Therin M, Doyle C (2001) Hydroxyapatite-coated acetabular components-histological and histomorphometric analysis of six cups retrieved at autopsy between three and seven years after successful implantation. J Bone Joint Surg 83:817–825. https://doi.org/10.1007/s00264-006-0226-4

    Article  CAS  PubMed  Google Scholar 

  35. Harman MK, Hodge WA, Banks SA (2003) Closed reduction of constrained total hip arthroplasty. Clin Orthop Relat Res 414:121–128. https://doi.org/10.1097/01.blo.0000073342.50837.4a

    Article  Google Scholar 

  36. Noble PC, Durrani SK, Usrey MM, Mathis KB, Bardakos NV (2012) Constrained cups appear incapable of meeting the demands of revision THA. Clin Orthop Relat Res 470:1907–1916. https://doi.org/10.1007/s11999-011-2212-4

    Article  PubMed  Google Scholar 

  37. Liu F, Jin ZM, Grigoris P, Hirt F, Rieker C (2003) Contact mechanics of metal-on-metal hip implants employing a metallic cup with a UHMWPE backing. J Instn Mech Engres 217:207–213. https://doi.org/10.1243/095441103765212703

    Article  CAS  Google Scholar 

  38. Uddin MS, Chan GWC (2018) Reducing stress concentration on the cup rim of hip implants under edge loading. Int J Numer Meth Biomed 35:e3149. https://doi.org/10.1002/cnm.3149

    Article  Google Scholar 

  39. Fuxing T, Mingzhe L, Zhongyi C, Xiangji L (2009) Formability analysis on the process of multi-point forming for titanium alloy retiary sheet. Int J Adv Manuf Technol 41:1059–1065. https://doi.org/10.1007/s00170-008-1548-8

    Article  Google Scholar 

  40. Jang YW, Lim D, Seo H, Lee MC, Lee OS, Lee YS (2018) Role of an anatomically contoured plate and metal block for balanced stability between the implant and lateral hinge in open-wedge high-tibial osteotomy. Arch Orthop Trauma Surg 138:911–920. https://doi.org/10.1007/s00402-018-2918-9

    Article  PubMed  Google Scholar 

  41. Bergstrom JS, Rimnac CM, Kurtz SM (2003) Prediction of multiaxial mechanical behavior for conventional and highly crosslinked UHMWPE using a hybrid constitutive model. Biomaterials 24:1365–1380. https://doi.org/10.1016/S0142-9612(02)00514-8

    Article  CAS  PubMed  Google Scholar 

  42. Guezmil M, Bensalah W, Mezlini S (2016) Tribological behavior of UHMWPE against TiAl6V4 and CoCr28Mo alloys under dry and lubricated conditions. J Mech Behav Biomed Mater 63:375–385. https://doi.org/10.1016/j.jmbbm.2016.07.002

    Article  CAS  PubMed  Google Scholar 

  43. Marino DJ, Ireifej SJ, Loughin CA (2012) Micro total hip replacement in dogs and cats. Vet Surg 41:121–129. https://doi.org/10.1111/j.1532-950X.2011.00933.x

    Article  PubMed  Google Scholar 

  44. Schiller TD (2017) BioMedtrix total hip replacement systems: an overview. Vet Clin Small Anim 47:899–916. https://doi.org/10.1016/j.cvsm.2017.03.005

    Article  Google Scholar 

  45. Dasari A, Misra RDK (2003) On the strain rate sensitivity of high density polyethylene and polypropylenes. Mater Sci Eng A 358:356–371. https://doi.org/10.1016/S0921-5093(03)00330-7

    Article  CAS  Google Scholar 

  46. Anderson MJ, Murray WR, Skinner HB (1994) Constrained acetabular components. J Arthroplast 9:17–23. https://doi.org/10.1016/0883-5403(94)90133-3

    Article  CAS  Google Scholar 

  47. Niinomi M (1998) Mechanical properties of biomedical titanium alloys. Mater Sci Eng A Struct Mater 243:231–236. https://doi.org/10.1016/S0921-5093(97)00806-X

    Article  Google Scholar 

  48. Brown EN, Willms RB, Gray GT, Rae PJ, Cady CM, Vecchio KS, Flowers J, Martinez MY (2007) Influence of molecular conformation on the constitutive response of polyethylene: a comparison of HDPE, UHMWPE, and PEX. Exp Mech 47:381–393. https://doi.org/10.1007/s11340-007-9045-9

    Article  CAS  Google Scholar 

  49. Muratoglu OK, Bragdon CR, O'Connor DO, Jasty M, Harris WH, Gul R, McGarry F (1999) Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials 20:1463–1470. https://doi.org/10.1016/S0142-9612(99)00039-3

    Article  CAS  PubMed  Google Scholar 

  50. Catelas I, Wimmer MA, Utzschneider S (2011) Polyethylene and metal wear particles: characteristics and biological effects. Semin Immunopathol 33:257–271. https://doi.org/10.1007/s00281-011-0242-3

    Article  CAS  PubMed  Google Scholar 

  51. Kobayashi S, Takaoka K, Tsukada A, Ueno M (1998) Polyethylene wear from femoral bipolar neck-cup impingement as a cause of femoral prosthetic loosening. Arch Orthop Trauma Surg 117:390–391. https://doi.org/10.1007/s004020050274

    Article  CAS  PubMed  Google Scholar 

  52. Shahemi N, Liza S, Abbas AA, Merican AM (2018) Long-term wear failure analysis of UHMWPE acetabular cup in total hip replacement. J Mech Behav Biomed Mater 87:1–9. https://doi.org/10.1016/j.jmbbm.2018.07.017

    Article  CAS  PubMed  Google Scholar 

  53. Padgett DE, Lipman J, Robie B, Nestor BJ (2006) Influence of total hip design on dislocation: a computer model and clinical analysis. Clin Orthop Relat Res 447:48–52. https://doi.org/10.1097/01.blo.0000218748.30236.40

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We express our deep thanks to Mr. Yasuto Shimizu, President of Shimizutech Corporation, for his warm encouragement for the present study. The authors acknowledge Mr. James A. Horvat, Kindai University, for carefully reading and giving critical comments on this manuscript’s language.

Funding

This work was supported in part by a grant of the Strategic Research Foundation at Private Universities from the Ministry of Education, Culture, Sports, Science, and Technology, Japan (MEXT), 2013–2017 (S1311045).

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization: Makoto Saito, Takayuki Kobayashi, and Ei Yamamoto;

Methodology: Yuki Kawamura, Yoshihito Matsufuji, Yoshinori Uwa, and Saiji Washio;

Formal analysis and investigation: Yuki Kawamura, Mitsushi Ohmasa, and Ei Yamamoto;

Writing—original draft preparation: Yuki Kawamura;

Writing—review and editing: Ei Yamamoto;

Funding acquisition: Ei Yamamoto;

Resources: Yoshihito Matsufuji and Makoto Saito;

Supervision: Takayuki Kobayashi and Ei Yamamoto;

All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ei Yamamoto.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kawamura, Y., Ohmasa, M., Kobayashi, T. et al. Estimation of the effects of inset heights and slit configurations in an acetabular cup on the pull-out behavior of an artificial hip joint with a structure for preventing dislocation using finite element analysis. Med Biol Eng Comput 58, 2587–2601 (2020). https://doi.org/10.1007/s11517-020-02231-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11517-020-02231-8

Keywords

Navigation