Skip to main content

Advertisement

Log in

Ultrasound-guided spinal injections: a feasibility study of a guidance system

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

Facet joint injections of analgesic agents are widely used to treat patients with lower back pain. The current standard-of-care for guiding the injection is fluoroscopy, which exposes the patient and physician to significant radiation. As an alternative, several ultrasound guidance systems have been proposed, but have not become the standard-of-care, mainly because of the difficulty in image interpretation by the anesthesiologist unfamiliar with the complex spinal sonography.

Methods

We introduce an ultrasound-based navigation system that allows for live 2D ultrasound images augmented with a patient-specific statistical model of the spine and relating this information to the position of the tracked injection needle. The model registration accuracy is assessed on ultrasound data obtained from nine subjects who had prior CT images as the gold standard for the statistical model. The clinical validity of our method is evaluated on four subjects (of an ongoing in vivo study) which underwent facet joint injections.

Results

The statistical model could be registered to the bone structures in the ultrasound volume with an average RMS accuracy of \(2.3\pm 0.4\) mm. The shape of the individual vertebrae could be estimated from the US volume with an average RMS surface distance error of \(1.5\pm 0.4\) mm. The facet joints could be identified by the statistical model with an average accuracy of \(5.1\pm 1.5\) mm.

Conclusions

The results of this initial feasibility assessment suggest that this ultrasound-based system is capable of providing information sufficient to guide facet joint injections. Further clinical studies are warranted.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

Notes

  1. http://perk-software.cs.queensu.ca/plus/doc/nightly/user/ApplicationfCal.html.

  2. http://www.assembla.com/spaces/plus/wiki/Printable_3D_models.

  3. http://perk-software.cs.queensu.ca/plus/doc/nightly/user/ProcedureVolumeReconstruction.html.

  4. www.itksnap.org.

References

  1. Boisvert J, Cheriet F, Pennec X, Labelle H, Ayache N (2008) Articulated spine models for 3-D reconstruction from partial radiographic data. IEEE Trans Biomed Eng 55(11):2565–2574

    Article  PubMed  Google Scholar 

  2. Boswell MV, Colson JD, Sehgal N, Dunbar EE, Epter R (2007) A systematic review of therapeutic facet joint interventions in chronic spinal pain. Pain Physician 10(1):229–253

    PubMed  Google Scholar 

  3. Chen ECS., Mousavi P, Gill S, Fichtinger G, Abolmaesumi P (2010) Ultrasound guided spine needle insertion. In: Proceedings of SPIE, vol 7625, pp 762538-1–762538-8

  4. Foroughi P, Boctor E, Swartz M, Taylor R, Fichtinger G (2007) Ultrasound bone segmentation using dynamic programming. IEEE Ultrason Symp 13(4):2523–2526

    Google Scholar 

  5. Greher M, Scharbert G, Kamolz LP, Beck H, Gustorff B, Kirchmair L, Kapral S (2004) Ultrasound-guided lumbar facet nerve block: a sonoanatomic study of a new methodologic approach. Anesthesiology 100(5):1242–1248

    Article  PubMed  Google Scholar 

  6. Khallaghi S, Mousavi P, Gong RH, Gill S, Boisvert J, Fichtinger G, Pichora D, Borschneck D, Abolmaesumi P (2010) Registration of a statistical shape model of the lumbar spine to 3D ultrasound images. In: MICCAI 2010, Part II, vol LNCS 6362, pp 68–75

  7. Lasso A, Heffter T, Rankin A, Pinter C, Ungi T, Fichtinger G (2014) PLUS: open-source toolkit for ultrasound-guided intervention systems. IEEE Trans Biomed Eng 61(10):2527–2537

  8. Malenfant PA. Gunka V, Beigi P, Rasoulian A, Rohling R, Dube A (2014) Accuracy of 3D ultrasound for identification of epidural needle skin insertion point in parturients; a prospective observational study. In: Society for Obstetric Anesthesia and Perinatology (SOAP) 46th annual meeting. Toronto, ON, Canada

  9. Moore J, Clarke C, Bainbridge D, Wedlake C, Wiles A, Pace D, Peters T (2009) Image guidance for spinal facet injections using tracked ultrasound. In: MICCAI 2009, Part I, vol LNCS 5761, pp 516–523

  10. Myronenko A, Song X (2010) Point set registration: coherent point drift. IEEE Trans Pattern Anal Mach Intell 32(12):2262–2275

    Article  PubMed  Google Scholar 

  11. Pennec X (2006) Intrinsic statistics on Riemannian manifolds: basic tools for geometric measurements. J Math Imaging Vis 25(1):127–154

    Article  Google Scholar 

  12. Pieper S, Halle M, Kikinis R (2004) 3D Slicer. In: IEEE international symposium on biomedical imaging: from nano to macro, pp 632–635

  13. Rasoulian A, Rohling R, Abolmaesumi P (2013) Augmentation of paramedian 3D ultrasound images of the spine. In: IPCAI 2013, vol LNCS 7915, pp 51–60

  14. Rasoulian A, Rohling R, Abolmaesumi P (2013) Lumbar spine segmentation using a statistical multi-vertebrae anatomical shape \(+\) pose model. IEEE Transactions on Medical Imaging 32(10):1890–1900

    Article  PubMed  Google Scholar 

  15. Roberts MG, Cootes TF, Pacheco E, Oh T, Adams JE (2009) Segmentation of lumbar vertebrae using part-based graphs and active appearance models. In: MICCAI, Part II, vol LNCS 5761, pp 1017–1024

  16. Rubin DI (2007) Epidemiology and risk factors for spine pain. Neurol Clin 25(2):353–371

    Article  PubMed  Google Scholar 

  17. Tokuda J, Fischer GS, Papademetris X, Yaniv Z, Ibanez L, Cheng P, Liu H, Blevins J, Arata J, Golby AJ, Kapur T, Pieper S, Burdette EC, Fichtinger G, Tempany CM, Hata N (2009) OpenIGTLink: an open network protocol for image-guided therapy environment. Int J Med Robot Comput Assist Surg 5(4):423–434

    Article  Google Scholar 

  18. Tran D, Kamani AA, Al-Attas E, Lessoway VA, Massey S, Rohling RN (2010) Single-operator real-time ultrasound-guidance to aim and insert a lumbar epidural needle. Can J Anesth 57(4):313–321

    Article  PubMed  Google Scholar 

  19. Ungi T, Abolmaesumi P, Jalal R, Welch M, Ayukawa I, Nagpal S, Lasso A, Jaeger M, Borschneck DP, Fichtinger G, Mousavi P (2012) Spinal needle navigation by tracked ultrasound snapshots. IEEE Trans Biomed Eng 59(10):2766–2772

    Article  PubMed  Google Scholar 

  20. Yan CX, Goulet B, Pelletier J, Chen SJS, Tampieri D, Collins DL (2011) Towards accurate, robust and practical ultrasound-CT registration of vertebrae for image-guided spine surgery. Int J Comput Assist Radiol Surg 6(4):523–537

    Article  PubMed  Google Scholar 

  21. Yoon SH, O’Brien SL, Tran M (2013) Ultrasound guided spine injections: Advancement over fluoroscopic guidance? Curr Phys Med Rehabil Rep 1(2):104–113

    Article  Google Scholar 

Download references

Acknowledgments

The authors would like to thank the Natural Sciences and Engineering Research Council (NSERC) and the Canadian Institutes of Health Research (CIHR) for supporting this research.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alexander Seitel.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Rasoulian, A., Seitel, A., Osborn, J. et al. Ultrasound-guided spinal injections: a feasibility study of a guidance system. Int J CARS 10, 1417–1425 (2015). https://doi.org/10.1007/s11548-015-1212-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-015-1212-3

Keywords

Navigation