Minimizing left ventricular stroke work with iterative learning flow profile control of rotary blood pumps

https://doi.org/10.1016/j.bspc.2016.09.001Get rights and content

Highlights

  • An iteratively learning flow controller for rotary blood pumps is proposed.

  • It is used to track flow profiles that minimize the left ventricular stroke work.

  • Feasibility demonstrated in hybrid mock circulatory loop experiments.

Abstract

Rotary blood pumps are gaining importance in the successful treatment of advanced heart failure. However, the application of fixed pump speeds is discussed controversially. Since the natural heart delivers pulsatile flow, many physicians presume that pulsatile pumping provides therapeutical advantages. To address this, we combine the technical advantages of continuous flow devices with the supposed physiological advantages of pulsatile flow. We present an iterative learning control (ILC) strategy for continuous flow ventricular assist devices that minimizes the left ventricular stroke work (LVSW). For that, a comprehensive nonlinear model for rotary blood pumps that is used for simulation and controller design is introduced. The controller is tested using a hardware-in-the-loop cardiovascular system simulator with a Medos deltastream DP1 blood pump. The tracking performance of the proposed ILC approach is compared to a benchmark controller that uses additional sensor information, both controllers significantly reduce the residual LVSW compared to the fixed speed case. In addition to decreasing ventricular load, the proposed ILC strategy can be used as an inner control loop to any physiological controller that sets reference flow profiles. The introduced controller might be useful for the investigation of effects of various pulsatile flow patterns independent from the type of VAD in future in vivo studies. The targeted manipulation of physiological quantities such as the residual cardiac work has the potential to considerably improve ventricular assist device therapy.

Introduction

Patients with advanced heart failure are unable to perform any physical activity without discomfort [1]. Approximately 1–2% of the adult population in developed countries suffers from heart failure [2]. In case of an unfavorable progression of the disease, heart transplantation remains the gold standard therapy. Due to the limited availability of donor organs and the rapid development of ventricular assist device (VAD) technology, the number of patients bridged to transplant or receiving destination therapy by implantation of a VAD increased significantly [3]. This trend is confirmed by the latest Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) report, which includes clinical information from 15,745 adult patients that received primary prospective implants between June 2006 and December 2014. The vast majority of these patients (84.4%) were treated with a left ventricular assist device (LVAD) [4].

The purpose of an LVAD is to relieve the native heart and to increase the cardiac output by pumping blood from the left ventricle into the aorta. In modern LVAD therapy, displacement pumps are virtually replaced by rotary blood pumps, as the latter are associated with dramatically lower overall adverse event rates [4]. These pumps are normally operated at a constant rotational speed; this leads to a reduced vascular pulsatility compared to physiological flow [5]. There is evidence that fixed speed support using continuous flow devices can cause aortic valve insufficiency [6], [7], gastrointestinal bleeding [8], and thromboembolic events [9]. It is not yet known whether these complications are induced by the pump type or the lack of pulsatile flow. A systematic overview of clinical studies comparing pulsatile and continuous flow support and their conflicted findings is given in [10]. These studies cannot separate pump design from flow effects as they are limited to the comparison of previous pulsatile technology to recent continuous flow devices. This issue may be addressed by pulsatile flow control of rotary blood pumps.

A control strategy that restores pulsatility might be the solution for combining the benefits of smaller and more reliable rotary blood pumps with the positive hemodynamic effects of pulsatile devices [11]. A recent review of studies investigating cardiac cycle-synchronized speed modulation of continuous flow devices [12] confirmed that such strategies can increase pulsatility in the systemic arterial circulation and enhance the unloading of the left ventricle. These studies focused on the modulation of pump speed as this is the control variable in standard VAD therapy and hence clinically available. However, the quantity that directly affects hemodynamics is the pump flow, which depends not only on the pump speed, but also on the differential pressure across the pump. Hence, it is preferable to directly control the pump flow [12]. It enhances the comparability of studies conducted with different VADs and heterogeneous groups of subjects. Ising et al. [13] demonstrated in silico that the arterial pulsatility and ventricular work can significantly be affected by the modulation of the pump flow. The specific control of VADs might eventually lead to higher myocardial recovery rates [14].

In this paper, we present an iteratively learning flow controller for rotary blood pumps. It enables good reference tracking of numerically optimal flow profiles that minimize the left ventricular stroke work (LVSW). Moreover, we show that pump flow modulation outperforms the fixed speed control operation case in terms of LVSW reduction, which is consistent with the literature. Therefore, we present a comprehensive nonlinear model for continuous-flow VADs. It is used to design and test the iterative learning control (ILC) strategy. ILC was chosen because it exploits the repetitive nature of the changing differential pressure across the pump to improve the flow tracking performance. Besides, it was already successfully applied in clinical trials in other areas of biomedical engineering [15], [16]. The improvement of the flow tracking performance is achieved by generating a control input trajectory that incorporates error information of previous heartbeats. The LVSW-optimal reference trajectories, which are solutions of a nonlinear program, allow a meaningful comparison of different control strategies. We tested two strategies: ILC combined with a PID feedback controller and a classical PID controller with PD disturbance feedforward control. Their performance was assessed in a hybrid mock circulatory loop with a cardiovascular system (CVS) model simulating severe heart failure.

This paper is structured as follows. Section 2 presents our approach and the test setup. In Section 3, in vitro results are analyzed and discussed. Our conclusions are drawn in the final section.

Section snippets

Materials and methods

This section begins by introducing a generic nonlinear model for rotary blood pumps. It is used to design and test the iteratively learning pump flow controller proposed in Section 2.2. The tracking performance of the ILC is tested with numerical optimal flow profiles described in Section 2.3 in a hybrid mock circulatory loop which is presented in Section 2.4. The last section gives a brief overview of the experimental procedure.

Results and discussion

Fig. 7 shows the tracking performance of the ILC approach in comparison to the benchmark PID controller for the two LVSW-optimal flow profiles. The pump flow is depicted for one cardiac cycle whose start is defined as the onset of the ventricular systole (usually marked by the R peak in ECG). The isovolumic relaxation of the ventricle approximately ends at 45% of the cardiac phase. From late systole to early systole of the following heartbeat, the differential pressure Δpd is negative. The left

Conclusions

Our study confirms that the appropriate pulsatile control of rotary blood pumps reduces the left ventricular stroke work. We presented a comprehensive model for rotary blood pumps that was used to design an iterative learning controller for the accurate tracking of optimal flow profiles. The tracking performance of the ILC approach is marginally better compared to the benchmark controller that utilizes additional sensors to measure the differential pressure across the pump. Importantly, the

References (40)

  • A. Mosterd et al.

    Clinical epidemiology of heart failure

    Heart

    (2007)
  • A. Cheng et al.

    Heart transplantation

    J. Thorac. Dis.

    (2014)
  • S.V. Deo et al.

    De novo aortic insufficiency during long-term support on a left ventricular assist device: a systematic review and meta-analysis

    ASAIO J.

    (2014)
  • H. Patel et al.

    Complications of continuous-flow mechanical circulatory support devices

    Clin. Med. Insights Cardiol.

    (2015)
  • K.G. Soucy et al.

    Rotary pumps and diminished pulsatility: do we need a pulse?

    ASAIO J.

    (2013)
  • A. Cheng et al.

    Comparison of continuous-flow and pulsatile-flow left ventricular assist devices: is there an advantage to pulsatility?

    Ann. Cardiothorac. Surg.

    (2014)
  • R. Amacher et al.

    Synchronized pulsatile speed control of turbodynamic left ventricular assist devices: review and prospects

    Artif. Organs

    (2014)
  • M. Ising et al.

    Flow modulation algorithms for continuous flow left ventricular assist devices to increase vascular pulsatility: a computer simulation study

    Cardiovasc. Eng. Technol.

    (2011)
  • C.T. Freeman et al.

    Iterative learning control in health care: Electrical stimulation and robotic-assisted upper-limb stroke rehabilitation

    IEEE Cont. Syst.

    (2012)
  • K.L. Meadmore et al.

    The application of precisely controlled functional electrical stimulation to the shoulder, elbow and wrist for upper limb stroke rehabilitation: a feasibility study.

    J. Neuroeng. Rehabil.

    (2014)
  • Cited by (23)

    • Feedback controller for restoring the basal hemodynamic condition with a rotary blood pump used as left ventricular assist device

      2020, Biomedical Signal Processing and Control
      Citation Excerpt :

      Feedback control systems have been proposed in the literature to fulfill physiological objectives using RBPs. In Bozkurt et al. [10] and Rüschen et al. [11] a physiological controller was proposed based on a RBP. In both works, the feedback variable chosen was the pump flow, but the choice of this variable requires invasive methods and expensive sensors for the flow measurements.

    • A suction index based control system for rotary blood pumps

      2020, Biomedical Signal Processing and Control
      Citation Excerpt :

      While these strategies detected ventricular suction after it happened, they cannot prevent suction and the resulting myocardial damage and arrhythmia risk. There have been a number of control algorithms that provide sufficient physiologic perfusion while simultaneously preventing suction [17–20]. However, these algorithms need implantation of pressure or flow sensors that may be unreliable for long-term use because of sensor drift or failure, thrombus formation, and onset of infection.

    • Estimation of left ventricular stroke work based on a large cohort of healthy children

      2020, Computers in Biology and Medicine
      Citation Excerpt :

      Especially in children, the standard values of SW for different ages, weights, and heights could improve clinical decision-making by monitoring cardiac function during the growth process [11]. From an engineering point of view, knowledge about the range of SW can be helpful to control mechanical circulatory support devices, such as ventricular assist devices (VAD) [12–14]. SW represents the energy provided by the ventricle to pump blood into the arteries.