An Evaluation of Triple Density Error Diffusion for Medical Monochrome LCDs

Nobutaka KUROKI
Nobuhiro OKA
Masahiro NUMA
Keisuke YAMAMOTO

Publication
IEICE TRANSACTIONS on Fundamentals of Electronics, Communications and Computer Sciences   Vol.E89-A    No.6    pp.1866-1868
Publication Date: 2006/06/01
Online ISSN: 1745-1337
DOI: 10.1093/ietfec/e89-a.6.1866
Print ISSN: 0916-8508
Type of Manuscript: LETTER
Category: Image
Keyword: 
error diffusion,  medical LCD,  sub-pixel modulation,  eye-model,  

Full Text: PDF(299.8KB)>>
Buy this Article



Summary: 
A triple density Error Diffusion for medical monochrome LCDs is proposed to improve their gray-scale precisions. In addition, a measurement technique of image qualities based on E-MSE (Eye model-based Mean Square Error) is proposed. Several conventional techniques for medical LCDs, such as Sub-Pixel Modulation and Error Diffusion, are evaluated based on E-MSE and the validity of the proposed technique is ensured objectively.