Loading [a11y]/accessibility-menu.js
Classical Preisach model of hysteretic behavior in a da Vinci instrument | IEEE Conference Publication | IEEE Xplore

Classical Preisach model of hysteretic behavior in a da Vinci instrument


Abstract:

Tendon-based instruments are widely used in both robotic and manual minimally invasive surgical procedures. Direct measurement of the interaction forces at the tip of suc...Show More

Abstract:

Tendon-based instruments are widely used in both robotic and manual minimally invasive surgical procedures. Direct measurement of the interaction forces at the tip of such instruments is difficult. As a result, methods for estimation of these forces are of substantial interest. In this paper, modeling of the input-output hysteretic behavior in a da Vinci instrument is addressed using the classical Preisach approach. The performance of the developed model is experimentally evaluated. The results obtained demonstrate that the classical Preisach model allows for sufficiently precise estimation of the forces at the tip of the da Vinci instrument.
Date of Conference: 12-15 July 2016
Date Added to IEEE Xplore: 29 September 2016
ISBN Information:
Conference Location: Banff, AB, Canada

Contact IEEE to Subscribe

References

References is not available for this document.