Automated temperature calculation method for DWI-thermometry: The usefulness of LV probability map on healthy subjects | IEEE Conference Publication | IEEE Xplore

Automated temperature calculation method for DWI-thermometry: The usefulness of LV probability map on healthy subjects


Abstract:

Diffusion-weight imaging (DWI) has already been incorporated as a regular sequence for patients. If DWI could indicate brain temperature without a complicated procedure, ...Show More

Abstract:

Diffusion-weight imaging (DWI) has already been incorporated as a regular sequence for patients. If DWI could indicate brain temperature without a complicated procedure, such information may greatly contribute to initial diagnosis. The temperature (T: °C) was calculated using the following equation form the diffusion coefficient (D): T= 2256.74/ln (4.39221/D) - 273.15. The cerebrospinal fluid region for automated temperature computation was segmented by lateral ventricle probability map which was constructed from 46 healthy volunteers. No significant differences were seen between temperatures using the proposed method and the manually segmented. The proposed method of fully automated deep brain temperature computation from DWI may prove feasible for application in MRI consoles.
Date of Conference: 03-07 July 2013
Date Added to IEEE Xplore: 26 September 2013
Electronic ISBN:978-1-4577-0216-7

ISSN Information:

PubMed ID: 24109733
Conference Location: Osaka, Japan

References

References is not available for this document.