Measuring chest circumference change during respiration with an electromagnetic biosensor | IEEE Conference Publication | IEEE Xplore

Measuring chest circumference change during respiration with an electromagnetic biosensor


Abstract:

In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneo...Show More

Abstract:

In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneously harvesting significant power. Chest circumference change can provide information on tidal volume, which is vital in assessing lung function. The chest circumference change is calculated from the motor's voltage output. Calculated values are within 0.95mm of measured circumference changes, with a standard deviation of 0.37mm. The wearable motor can also harvest at least 29.4μW during normal breathing.
Date of Conference: 03-07 July 2013
Date Added to IEEE Xplore: 26 September 2013
Electronic ISBN:978-1-4577-0216-7

ISSN Information:

PubMed ID: 24110093
Conference Location: Osaka, Japan

Contact IEEE to Subscribe

References

References is not available for this document.