Abstract:
In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneo...Show MoreMetadata
Abstract:
In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneously harvesting significant power. Chest circumference change can provide information on tidal volume, which is vital in assessing lung function. The chest circumference change is calculated from the motor's voltage output. Calculated values are within 0.95mm of measured circumference changes, with a standard deviation of 0.37mm. The wearable motor can also harvest at least 29.4μW during normal breathing.
Published in: 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 03-07 July 2013
Date Added to IEEE Xplore: 26 September 2013
Electronic ISBN:978-1-4577-0216-7
ISSN Information:
PubMed ID: 24110093