Abstract:
The clinical demand for technologies to monitor Blood Pressure (BP) in ambulatory scenarios with minimal use of inflation cuffs is strong: new generation of BP monitors a...Show MoreMetadata
Abstract:
The clinical demand for technologies to monitor Blood Pressure (BP) in ambulatory scenarios with minimal use of inflation cuffs is strong: new generation of BP monitors are expected to be not only accurate, but also non-occlusive. In this paper we review recent advances on the use of the so-called Pulse Wave Velocity (PWV) technologies to estimate BP in a beat-by-beat basis. After introducing the working principle and underlying methodological limitations, two implementation examples are provided. Pilot studies have demonstrated that novel PWV-based BP monitors depict accuracy scores falling within the limits of the British Hypertensive Society (BHS) Grade A standard. The reported techniques pave the way towards ambulatory-compliant, continuous and non-occlusive BP monitoring devices, where the use of inflation cuffs is drastically reduced.