Abstract:

Monte Carlo estimation of the giant-dipole-resonance (GRN) photoneutrons inside the Elekta Precise LINAC head (emitting a 15 MV photon beam) were performed using the MCNP...Show More

Abstract:

Monte Carlo estimation of the giant-dipole-resonance (GRN) photoneutrons inside the Elekta Precise LINAC head (emitting a 15 MV photon beam) were performed using the MCNP6 (general-purpose Monte Carlo N-Particle code, version 6). Each component of LINAC head geometry and materials were modelled in detail using the given manufacturer information. Primary photons generate photoneutrons and its transport across the treatment head was simulated, including the (n, γ) reactions which undergo activation products. The MCNP6 was used to develop a method for quantifying the activation of accelerator components. The approach described in this paper is useful in quantifying the origin and the amount of nuclear activation.