Abstract:
In the last century, medicine showed considerable advancements in terms of new technologies, devices and diagnostic/therapeutic strategies. Those advantages led to a sign...Show MoreMetadata
Abstract:
In the last century, medicine showed considerable advancements in terms of new technologies, devices and diagnostic/therapeutic strategies. Those advantages led to a significant reduction of invasiveness and an improvement of surgical outcomes. In this framework, a computer-assisted surgical robotic platform able to perform non-invasive Focused Ultrasound Surgery (FUS) - the FUTURA platform - has the ambitious goal to improve accuracy, safety and flexibility of the treatment, with respect to current FUS procedures. Aim of this work is to present the current implementation of the robotic platform and the preliminary results about high intensity focused ultrasound (HIFU) delivery in in-vitro conditions, under 3D ultrasound identification and monitoring. Tests demonstrated that the average accuracy of the HIFU delivery is lower than 0.7 mm in both X and Y radial directions and 3.7 mm in the axial direction (Z) with respect to the HIFU transducer active surface.
Published in: 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)