Abstract:
Inductive powering for implanted medical devices is a commonly employed technique, that allows for implants to avoid more dangerous methods such as the use of transcutane...Show MoreMetadata
Abstract:
Inductive powering for implanted medical devices is a commonly employed technique, that allows for implants to avoid more dangerous methods such as the use of transcutaneous wires or implanted batteries. However, wireless powering in this way also comes with a number of difficulties and conflicting requirements, which are often met by using designs based on compromise. In particular, one aspect common to most inductive power links is that they are driven with a fixed frequency, which may not be optimal depending on factors such as coupling and load. In this paper, a method is proposed in which an inductive power link is driven by a frequency that is maintained at an optimum value fopt , to ensure that the link is in resonance. In order to maintain this resonance, a phase tracking technique is employed at the primary side of the link; this allows for compensation of changes in coil separation and load. The technique is shown to provide significant improvements in maintained secondary voltage and efficiency for a range of loads when the link is overcoupled.
Published in: 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 25-29 August 2015
Date Added to IEEE Xplore: 05 November 2015
ISBN Information:
ISSN Information:
PubMed ID: 26737530