Abstract:
Electrosurgery is a commonly used device in the operating room, but it has some adverse effects, which are only partly described in literature. Interference issues are we...Show MoreMetadata
Abstract:
Electrosurgery is a commonly used device in the operating room, but it has some adverse effects, which are only partly described in literature. Interference issues are well described, but unintended heating in and around implants is not well studied. We simulated different scenarios using a Finite Element Model to investigate unintended heating caused by electrosurgery. We looked at different shapes, sizes, and active electrode placements. We found that all these factors play a role in the amount of heating.
Published in: 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 25-29 August 2015
Date Added to IEEE Xplore: 05 November 2015
ISBN Information:
ISSN Information:
PubMed ID: 26737611