Abstract:
This paper presents a low-power front-end IC for wearable health monitoring systems. The IC, designed in a standard 0.13μm CMOS technology, fully integrates a low-noise a...Show MoreMetadata
Abstract:
This paper presents a low-power front-end IC for wearable health monitoring systems. The IC, designed in a standard 0.13μm CMOS technology, fully integrates a low-noise analog front-end (AFE) to process the weak bio-signals, followed by an analog-to-digital converter (ADC) to digitize the extracted signals. An AC-coupled driving buffer, that interfaces between the AFE and the ADC is introduced to scale down the power supply of the ADC. The power consumption decreases by 50% compared to the case without power supply scaling. The AFE passes signals from 0.5Hz to 280Hz and from 0.7Hz to 160Hz with a simulated input referred noise of 1.6μVrms and achieves a maximum gain of 35dB/41dB respectively, with a noise-efficiency factor (NEF) of the AFE is 1. The 8-bit ADC achieves a simulated 7.96-bit resolution at 10KS/s sampling rate under 0.5V supply voltage. The overall system consumes only 0.86μW at dual supply voltages of 1V (AFE) and 0.5 V (ADC).
Published in: 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 16-20 August 2016
Date Added to IEEE Xplore: 18 October 2016
ISBN Information:
ISSN Information:
PubMed ID: 28268699