The measurement of intraocular pressure (IOP) is an important vital sign for the eye, particularly for the diagnosis of glaucoma. Procedures for measuring IOP have been u...Show MoreMetadata
Abstract:
The measurement of intraocular pressure (IOP) is an important vital sign for the eye, particularly for the diagnosis of glaucoma. Procedures for measuring IOP have been used by eye care professionals for over 100 years, but those without access to such professionals often go undiagnosed. We present a smartphone-based system that can be operated by minimally trained users to measure IOP. The system emulates fixed-force tonometry using a low-cost mechanical attachment to the smartphone. Video is captured through the attachment and then processed in real-time to provide an absolute estimate of the patient's intraocular pressure. Our preliminary assessment with two ex vivo porcine eyes demonstrates that the system follows a baseline physical model with correlations of 0.89 and 0.88.
Published in: 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 16-20 August 2016
Date Added to IEEE Xplore: 18 October 2016
ISBN Information:
ISSN Information: