Abstract:
This paper proposes design considerations that need to be followed in order to eliminate potential sources of artefact that could distort a recorded neural signal. The ar...Show MoreMetadata
Abstract:
This paper proposes design considerations that need to be followed in order to eliminate potential sources of artefact that could distort a recorded neural signal. The artefact that appears in a recorded signal has a combination of potential sources each of which contributes towards its formation. As such, these sources of artefact have been addressed in three main categories: a) electronics artefact, b) encapsulation artefact and c) interface artefact. Each source (component) is analyzed further and appropriate design techniques and considerations are suggested towards its mitigation.
Published in: 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 23-27 July 2019
Date Added to IEEE Xplore: 07 October 2019
ISBN Information:
ISSN Information:
PubMed ID: 31946688