Loading [a11y]/accessibility-menu.js
Reconstruction with angular compensation in respiratory-gated cardiac SPECT | IEEE Conference Publication | IEEE Xplore

Reconstruction with angular compensation in respiratory-gated cardiac SPECT


Abstract:

In respiratory-gated cardiac SPECT with amplitude binning, the acquisition time can vary greatly both among respiratory gates and among acquisition angles within each gat...Show More

Abstract:

In respiratory-gated cardiac SPECT with amplitude binning, the acquisition time can vary greatly both among respiratory gates and among acquisition angles within each gate. If not properly accounted for, this uneven distribution in acquired data statistics will lead to limited-angle artifacts in reconstruction, which in turn can impact on the accuracy of respiratory motion correction. We investigate a compensation scheme for this uneven distribution by directly taking into account in the imaging model the actual acquisition time at different projection angles. In the experiment, we demonstrated this approach with simulated NCAT imaging data, for which quantitative results were obtained on both the reconstructed myocardium and estimated motion; we also tested the proposed approach on a set of clinical acquisition. The results show that the proposed approach could effectively suppress the limited-angle artifacts and improve the reconstruction in terms of both image accuracy and lesion detectability.
Date of Conference: 27-30 October 2014
Date Added to IEEE Xplore: 29 January 2015
Electronic ISBN:978-1-4799-5751-4

ISSN Information:

Conference Location: Paris, France

Contact IEEE to Subscribe

References

References is not available for this document.