Loading [a11y]/accessibility-menu.js
Design and Test of an In-Vivo Robotic Camera Integrated with Optimized Illumination System for Single-port Laparoscopic Surgery | IEEE Conference Publication | IEEE Xplore

Design and Test of an In-Vivo Robotic Camera Integrated with Optimized Illumination System for Single-port Laparoscopic Surgery


Abstract:

This paper proposes a novel in-vivo robotic laparo-scopic camera design with an optimized illumination system, which is a crucial component for achieving high imaging qua...Show More

Abstract:

This paper proposes a novel in-vivo robotic laparo-scopic camera design with an optimized illumination system, which is a crucial component for achieving high imaging quality. The robotic camera design with three extendable wings can reserve sufficient on-board space to harbor the optimized illumination system without affecting the compactness of the camera. We contribute a freeform optical lens design method and develop three miniature optical lenses for the LEDs to achieve greater than 95% illumination uniformity, greater than 14, 000 lx illuminance on a target plane with a distance of 100 mm, and greater than 89% optical efficiency. The prototype is implemented and experimentally tested, which demonstrates great performance of the in-vivo robotic laparoscopic camera and the significance of the optimized illumination system.
Date of Conference: 21-25 May 2018
Date Added to IEEE Xplore: 13 September 2018
ISBN Information:
Electronic ISSN: 2577-087X
Conference Location: Brisbane, QLD, Australia

Contact IEEE to Subscribe

References

References is not available for this document.