Abstract:
This paper describes a mixed-signal ECG processing platform with an 12-bit ADC architecture that can adapt its sampling rate according to the input signals rate of change...Show MoreMetadata
Abstract:
This paper describes a mixed-signal ECG processing platform with an 12-bit ADC architecture that can adapt its sampling rate according to the input signals rate of change. This enables the sampling of ECG signals with significantly reduced data rate without loss of information. The presented adaptive sampling scheme reduces the ADC power consumption, enables the processing of ECG signals with lower power consumption, and reduces the power consumption of the radio while streaming the ECG signals. The test results show that running a CWT-based R peak detection algorithm using the adaptively sampled ECG signals consumes only 45.6μW and it leads to 36% less overall system power consumption.
Published in: 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society
Date of Conference: 30 August 2011 - 03 September 2011
Date Added to IEEE Xplore: 01 December 2011
ISBN Information:
ISSN Information:
PubMed ID: 22254775