MRI inter-packet movement correction for images acquired with non-complementary data | IEEE Conference Publication | IEEE Xplore

MRI inter-packet movement correction for images acquired with non-complementary data


Abstract:

Movement during the acquisition of magnetic resonance images can cause artifacts that interfere with subsequent image analysis. In this paper we address the problem of in...Show More

Abstract:

Movement during the acquisition of magnetic resonance images can cause artifacts that interfere with subsequent image analysis. In this paper we address the problem of inter-packet motion and provide a method to minimize errors associated with this artifact. The procedure is based on an iterative packet-to-volume registration process and does not require complementary information such as multimodal acquisitions or protocols that provide redundant volume data. A Kaiser-Bessel function is used to interpolate missing data. Experiments with simulated data demonstrate that the packet-to-volume registration improves greatly after a single iteration and maintains improvement for the following iterations, while experiments with real data demonstrate a substantial reduction in associated artifacts and improvement in quality. In both cases anatomical integrity is preserved after reconstruction.
Date of Conference: 14-17 May 2008
Date Added to IEEE Xplore: 13 June 2008
ISBN Information:

ISSN Information:

Conference Location: Paris, France

Contact IEEE to Subscribe

References

References is not available for this document.