Abstract:
The aim of this study is to investigate the influence of stent axial orientation on saccular aneurysm hemodynamics. Two commercial stents, Neuroform stent (stent 1) and Z...Show MoreMetadata
Abstract:
The aim of this study is to investigate the influence of stent axial orientation on saccular aneurysm hemodynamics. Two commercial stents, Neuroform stent (stent 1) and Zilver stent (stent 2) are modeled in this study. Both stents are virtually deployed with four different axial orientations to fit into the luminal surface of a patient-specific internal carotid artery (ICA) aneurysm model. Computational hemodynamic analyses are carried out in unstented and stented aneurysm models. The intra-aneurismal flows of stented aneurysm models show disturbed and complex flow patterns while the flow activities and the forces acting on the aneurysm wall are generally alleviated by stenting. The influence of the axial orientation of the stent on the aneurysm hemodynamics is more significant for stent 2 which has a larger strut size compared to stent 1. Interestingly, the flow activity in the aneurysm is rather increased when the intra-luminal scaffolding of stent is not sufficient.
Date of Conference: 28 June 2009 - 01 July 2009
Date Added to IEEE Xplore: 07 August 2009
ISBN Information: