Loading [MathJax]/extensions/TeX/ieee_stixext.js
Automatic blood pool identification in contrast ultrasound using principal component analysis | IEEE Conference Publication | IEEE Xplore
Scheduled Maintenance: On Monday, 27 January, the IEEE Xplore Author Profile management portal will undergo scheduled maintenance from 9:00-11:00 AM ET (1400-1600 UTC). During this time, access to the portal will be unavailable. We apologize for any inconvenience.

Automatic blood pool identification in contrast ultrasound using principal component analysis


Abstract:

Several cardiovascular parameters of clinical interest can be assessed by indicator dilution techniques. Ultrasound contrast agents have been proposed as non-invasive ind...Show More

Abstract:

Several cardiovascular parameters of clinical interest can be assessed by indicator dilution techniques. Ultrasound contrast agents have been proposed as non-invasive indicator, showing promising results for blood volume estimation. However, the definition of an optimal region of interest for quantification of the indicator remains a critical step in the procedure, usually performed manually. In this work we present an automatic method to extract indicator dilution curves. Dimensionality reduction is achieved by principal component analysis followed by clustering to identify the different regions of interest. The method is evaluated on in vitro and in vivo datasets, compared to manually defined regions. The average difference was −3.47% ± 3.58% for in vitro volume estimates and the error was 1.29% ± 2.54% for trans-pulmonary mean transit time estimation. The new method allows kinetic parameter estimates in close agreement with those obtained manually; therefore it is a promising alternative for indicator dilution curve extraction.
Date of Conference: 29 April 2014 - 02 May 2014
Date Added to IEEE Xplore: 31 July 2014
Electronic ISBN:978-1-4673-1961-4

ISSN Information:

Conference Location: Beijing, China

References

References is not available for this document.