Abstract:
We describe a capacitance sensor array that has been incorporated into a lab-on-CMOS system for applications in monitoring cell viability. This paper presents measured ex...Show MoreMetadata
Abstract:
We describe a capacitance sensor array that has been incorporated into a lab-on-CMOS system for applications in monitoring cell viability. This paper presents measured experimental results of a high dynamic range capacitance sensor array. The sensors have been characterized and the data fitted to analytical models, allowing end-to-end calibration of the responses. The sensor exhibits a sensitivity of up to 593 kHz/fF, resolution down to 17.5 aF, and an input sensing range of 12 fF.
Date of Conference: 28-31 May 2017
Date Added to IEEE Xplore: 28 September 2017
ISBN Information:
Electronic ISSN: 2379-447X