Loading web-font TeX/Main/Regular
A CMOS Electrochemical Biochip With 32-32 Three-Electrode Voltammetry Pixels | IEEE Journals & Magazine | IEEE Xplore

A CMOS Electrochemical Biochip With 32 \times 32 Three-Electrode Voltammetry Pixels


Abstract:

A CMOS-integrated electrochemical biochip for high-performance molecular testing is presented. The system includes an array of 32×32 three-electrode voltammetry pixels an...Show More

Abstract:

A CMOS-integrated electrochemical biochip for high-performance molecular testing is presented. The system includes an array of 32×32 three-electrode voltammetry pixels and on-chip temperature control between 25 and 95 °C. Each 100 μm × 100 μm pixel includes a CMOS-compatible and chemo-stable amorphous carbon electrode transducer connected to in-pixel ΣA current detection circuitry with 280 fA rms input-referred noise (0.1-20 Hz bandwidth) and 93 dB dynamic range (DR). Array-based DNA detection assays are implemented, and successful DNA melt-analysis and real-time label-free DNA hybridization detection are reported.
Published in: IEEE Journal of Solid-State Circuits ( Volume: 54, Issue: 11, November 2019)
Page(s): 2980 - 2990
Date of Publication: 15 October 2019

ISSN Information:


References

References is not available for this document.