Loading [a11y]/accessibility-menu.js
PTT based continuous time non-invasive blood pressure system | IEEE Conference Publication | IEEE Xplore

PTT based continuous time non-invasive blood pressure system


Abstract:

This paper introduces the design and implementation of non-invasive, continuous time, cuff-less, blood pressure (BP) system based on pulse transmission time delay (PTT). ...Show More

Abstract:

This paper introduces the design and implementation of non-invasive, continuous time, cuff-less, blood pressure (BP) system based on pulse transmission time delay (PTT). The wearable BP device processing and monitoring both are performed using Smartphone. The PPG sensor and the integrated ECG sensor are wireless connected to smartphone. An adaptive threshold peak detection algorithm is proposed and tested to enhance the measurement accuracy. The proposed BP system achieves a systolic BP (SBP):- 1.36±7.51 mmHg and diastolic BP (DBP):- 2.44±3.49 mmHg. A new 4-LED PPG sensor is also introduced which can save 30% power consumption and reduce the interference of motion artifact (MA).
Date of Conference: 16-19 October 2016
Date Added to IEEE Xplore: 06 March 2017
ISBN Information:
Electronic ISSN: 1558-3899
Conference Location: Abu Dhabi

Contact IEEE to Subscribe

References

References is not available for this document.