Implantable Stimulator for Epileptic Seizure Suppression With Loading Impedance Adaptability


Abstract:

The implantable stimulator for epileptic seizure suppression with loading impedance adaptability was proposed in this work. The stimulator consisted of the high voltage g...Show More

Abstract:

The implantable stimulator for epileptic seizure suppression with loading impedance adaptability was proposed in this work. The stimulator consisted of the high voltage generator, output driver, adaptor, and switches, can constantly provide the required 40-\mu{\rm A} stimulus currents, as the loading impedance varied within 10 {\rm k}\Omega –300 {\rm k}\Omega. The performances of this design have been successfully verified in silicon chip fabricated by a 0.35- \mu{\rm m} 3.3-V/24-V CMOS process. The power consumption of this work was only 1.1 mW–1.4 mW. The animal test results with the fabricated chip of proposed design have successfully verified in the Long-Evans rats with epileptic seizures.
Published in: IEEE Transactions on Biomedical Circuits and Systems ( Volume: 7, Issue: 2, April 2013)
Page(s): 196 - 203
Date of Publication: 03 July 2012

ISSN Information:

PubMed ID: 23853302