A Dedicated Two-Channel Phased-Array Receiver Coil for High-Resolution MRI of the Rat Knee Cartilage at 7 T


Abstract:

In the field of small animal studies, the array coil imaging has become increasingly important. In this paper, a dedicated two-channel array coil operating at 300 MHz (7 ...Show More

Abstract:

In the field of small animal studies, the array coil imaging has become increasingly important. In this paper, a dedicated two-channel array coil operating at 300 MHz (7 T) for high-resolution MRI (HR-MRI) of the rat knee cartilage is presented. The average gain in signal-to-noise ratio (SNR) compared to a 15-mm multipurpose surface coil was 2.2. This SNR gain was used to improve the spatial resolution of 3-D acquisitions by decreasing the voxel size from 59 times 59 times 156 mum3 to 51 times 51 times 94 mum3 without time penalty. Also, a set of two array coils was used to perform a simultaneous acquisition of both knee joints of a rat, maintaining the same scanning time without SNR or spatial resolution degradation compared to the single knee joint acquisition. This two-channel array coil is a key element to perform HR-MRI and extract cartilage morphological parameters such as thickness and volume.
Published in: IEEE Transactions on Biomedical Engineering ( Volume: 56, Issue: 12, December 2009)
Page(s): 2891 - 2897
Date of Publication: 03 October 2008

ISSN Information:

PubMed ID: 19932985