Loading [a11y]/accessibility-menu.js
A Novel Wireless Photoplethysmography Blood-Flow Volume Sensor for Assessing Arteriovenous Fistula of Hemodialysis Patients | IEEE Journals & Magazine | IEEE Xplore

A Novel Wireless Photoplethysmography Blood-Flow Volume Sensor for Assessing Arteriovenous Fistula of Hemodialysis Patients


Abstract:

A novel blood-flow volume (BFV) sensor is presented for assessing quality of arteriovenous fistula in hemodialysis patients via noninvasive reflectance-type photoplethysm...Show More

Abstract:

A novel blood-flow volume (BFV) sensor is presented for assessing quality of arteriovenous fistula in hemodialysis patients via noninvasive reflectance-type photoplethysmography (PPG). BFV is nowadays in clinic practices evaluated using an ultrasound Doppler monitor, which is expensive, bulky, and can only be operated by well-trained medical personnels. This study is devoted to develop a low-cost, small-sized, portable, and easy-to-use PPG sensor that is capable of continuous measurement of BFV. New designs of front-end analog circuits, signal processing, and an intelligent neural network calibration method are employed to finally achieve high correlation (R2 = 0.7176), as opposed to the ultrasound Doppler monitor, with the root-mean-squared errors successfully controlled under 289 ml/min.
Published in: IEEE Transactions on Industrial Electronics ( Volume: 64, Issue: 12, December 2017)
Page(s): 9626 - 9635
Date of Publication: 08 May 2017