Loading [a11y]/accessibility-menu.js
Phase Comparison of High-Current Shunts up to 100 kHz | IEEE Journals & Magazine | IEEE Xplore

Phase Comparison of High-Current Shunts up to 100 kHz


Abstract:

A comparison of the phase error of existing shunts for currents between 10 and 100 A and frequencies from 500 Hz to 100 kHz has been organized. Measurement processes were...Show More

Abstract:

A comparison of the phase error of existing shunts for currents between 10 and 100 A and frequencies from 500 Hz to 100 kHz has been organized. Measurement processes were performed at Istituto Nationale di Ricerca Metrologica on some shunts using the first prototype of a new type of a phase comparator and a step-up method for data processing. Results of the relative measurement processes have shown good repeatability. The method for evaluation of reference values was based on similar shunts and suitable assumptions and derived by extrapolation. For the type of resistance elements employed in these shunts, the derivation of the reference is not very accurate and still needs to be improved.
Published in: IEEE Transactions on Instrumentation and Measurement ( Volume: 60, Issue: 7, July 2011)
Page(s): 2359 - 2365
Date of Publication: 10 March 2011

ISSN Information:


Contact IEEE to Subscribe

References

References is not available for this document.