Loading [a11y]/accessibility-menu.js
Optimized Reconstruction Algorithm for Helical CT With Fractional Pitch Between 1PI and 3PI | IEEE Journals & Magazine | IEEE Xplore

Optimized Reconstruction Algorithm for Helical CT With Fractional Pitch Between 1PI and 3PI


Abstract:

We propose an approximate approach to use redundant data outside the 1PI window within the exact Katsevich reconstruction framework. The proposed algorithm allows a flexi...Show More

Abstract:

We propose an approximate approach to use redundant data outside the 1PI window within the exact Katsevich reconstruction framework. The proposed algorithm allows a flexible selection of the helical pitch, which is useful for clinical applications. Our idea is an extension of the one proposed by Kohler, Bontus, and Koken (2006). It is based on optimizing the contribution weights of convolution families used in exact Katsevich 3PI algorithms, so that the total weight of each Radon plane is as close to 1 as possible. Optimization is based on solving a least squares problem subject to linear constrains. Numerical evaluation shows good noise and artifact reduction properties of the proposed algorithm.
Published in: IEEE Transactions on Medical Imaging ( Volume: 28, Issue: 7, July 2009)
Page(s): 982 - 990
Date of Publication: 10 February 2009

ISSN Information:

PubMed ID: 19211349