25 September 2014 Vicarious calibration of Terra/ASTER/VNIR with desert scenes together with cross calibration
Author Affiliations +
Abstract
Fourteen and half years of vicarious calibration and cross calibration for Terra/ASTER/VNIR (Visible and Near Infrared Radiometer, one of three ASTER mission instruments onboard Terra satellite) are summarized together with an error budget analysis. As the results, it is found that Onboard Calibration (OBC) data derived Radiometric Calibration Coefficient (RCC) is situated within a range of uncertainty of vicarious calibration except the period after approximately 4000 days after launch. In other word, uncertainty of OBC RCC is overlapped with uncertainty of vicarious calibration for about 4000 days since launch. Meanwhile, cross calibration with MODIS, MISR and ETM+ supports vicarious calibration of VNIR. In other word, Cross RCC shows much consistent trend to Vicarious RCC than OBC RCC.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
K. Arai "Vicarious calibration of Terra/ASTER/VNIR with desert scenes together with cross calibration", Proc. SPIE 9218, Earth Observing Systems XIX, 92180R (25 September 2014); https://doi.org/10.1117/12.2060226
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Error analysis

Reflectivity

MODIS

Atmospheric optics

Optical calibration

Satellites

RELATED CONTENT

Radiometric cross-calibration of Terra ASTER and MODIS
Proceedings of SPIE (September 08 2015)
Landsat-7 ETM+ radiometric calibration status
Proceedings of SPIE (September 19 2016)
Utility of MODIS-Terra for ocean color applications
Proceedings of SPIE (October 05 2007)
A comparison of the Antarctic Dome C and Sonoran Desert...
Proceedings of SPIE (August 12 2010)

Back to Top