Paper
24 March 2016 Validated novel software to measure the conspicuity index of lesions in DICOM images
Author Affiliations +
Abstract
A novel software programme and associated Excel spreadsheet has been developed to provide an objective measure of the expected visual detectability of focal abnormalities within DICOM images. ROIs are drawn around the abnormality, the software then fits the lesion using a least squares method to recognize the edges of the lesion based on the full width half maximum. 180 line profiles are then plotted around the lesion, giving 360 edge profiles.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
K. R. Szczepura and D. J. Manning "Validated novel software to measure the conspicuity index of lesions in DICOM images", Proc. SPIE 9787, Medical Imaging 2016: Image Perception, Observer Performance, and Technology Assessment, 978703 (24 March 2016); https://doi.org/10.1117/12.2216211