8 March 2019 Dual-modality phantom for evaluating x-ray/echo registration accuracy
Lindsay E. Bodart, Timothy J. Hall, Jacob K. Ellis, Benjamin R. Ciske, Martin Wagner, Amish N. Raval, Michael A. Speidel
Author Affiliations +
Abstract
Transcatheter interventions for structural heart disease demand real-time visualization of catheter devices and their relationship to cardiac anatomy. Co-registration of x-ray fluoroscopy with echocardiography has been proposed to provide the necessary device and soft tissue visualization for these procedures. Development of real-time 3D x-ray/echo registration systems with device tracking has been hampered by the lack of a suitable test model. This study presents a phantom that is compatible with x-ray, CT, transthoracic (TTE), and transesophageal echo (TEE) for testing the feasibility and accuracy of new registration solutions. The phantom consists of a 20.3-cm diameter, 15-cm tall cylindrical shell with acoustic windows for TTE and an access port for a TEE probe. The interior contains 24 dual-modality targets, 5-mm in diameter, suspended in a three-turn helix occupying a volume that is similar to an adult heart. An ultrasound-compatible, tissue-mimicking slurry medium fills the remainder of the phantom. The dual-modality targets are agar based with barium sulfate (BaSO4) powder and glass beads added to generate contrast in both x-ray and ultrasound. Appropriate concentrations of these additives were determined experimentally with contrast measurements in x-ray, CT, and ultrasound. Selected concentrations were 150 mg/mL BaSO4 and 100 mg/mL of 53-63 μm diameter glass beads. Average target contrast (± SD) was 16% ± 2% in x-ray fluoroscopy (90 kV) and 1805 ± 99 HU in CT (100 kV). In ultrasound, target CNR was 4.30 ± 0.62 in 2D B-mode and 4.03 ±1.06 in 4D-mode images acquired at a center frequency of 2.8 MHz.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lindsay E. Bodart, Timothy J. Hall, Jacob K. Ellis, Benjamin R. Ciske, Martin Wagner, Amish N. Raval, and Michael A. Speidel "Dual-modality phantom for evaluating x-ray/echo registration accuracy", Proc. SPIE 10951, Medical Imaging 2019: Image-Guided Procedures, Robotic Interventions, and Modeling, 109511V (8 March 2019); https://doi.org/10.1117/12.2512908
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Image registration

Ultrasonography

Fluoroscopy

Heart

Visualization

X-ray imaging

Back to Top