IEICE Transactions on Electronics
Online ISSN : 1745-1353
Print ISSN : 0916-8524
Special Section on Microwave and Millimeter-Wave Technologies
Double Step Technique for Accurate Microwave High Attenuation Measurements
Anton WIDARTA
Author information
JOURNAL FREE ACCESS

2024 Volume E107.C Issue 10 Pages 349-354

Details
Abstract

A double step attenuation measurement technique using a non-isolating gauge block attenuator (GBA) has been proposed for accurate measurements of radio frequency and microwave high attenuation. For fixed attenuator as a device under test (DUT), a medium value (≤60dB) attenuator is used as the GBA which connected directly between the test ports, then high attenuation of the DUT is measured in two setups as follows. 1) Thru and GBA with normal power level and 2) GBA and DUT with higher power level. This approach removes the need to isolate the GBA, therefore, accurate measurements of high attenuation can be obtained simply over a broad frequency range. For variable or step attenuator as a DUT, one of the attenuation sections of the DUT is applied as the GBA. Detailed analyses and those verification measurements are carried out both for fixed attenuator, as well as for variable attenuator and show good agreement.

References (31)
Content from these authors
© 2024 The Institute of Electronics, Information and Communication Engineers
Previous article Next article
feedback