IEICE Transactions on Communications
Online ISSN : 1745-1345
Print ISSN : 0916-8516
Regular Section
Experimental Evaluation of SAR around an Implanted Cardiac Pacemaker Caused by Mobile Radio Terminal
Yuta ENDOKazuyuki SAITOSoichi WATANABEMasaharu TAKAHASHIKoichi ITO
Author information
JOURNAL RESTRICTED ACCESS

2012 Volume E95.B Issue 6 Pages 2129-2132

Details
Abstract

Although the effect of electromagnetic interference on an implanted cardiac pacemaker due to a nearby mobile phone has been investigated, there have been few studies on the enhancement of the specific absorption rate (SAR) around an implanted cardiac pacemaker due to a nearby mobile phone. In this study, the SAR distribution around a pacemaker model embedded in a parallelepiped torso phantom when a mobile phone was nearby was numerically calculated and experimentally measured. The results of both investigations showed a characteristic SAR distribution. The system presented can be used to estimate the effects of electromagnetic interference on implanted electric circuits and thus could lead to the development of guidelines for the safe use of mobile radio terminals near people with medical implants.