single-au.php

IJAT Vol.5 No.2 pp. 115-119
doi: 10.20965/ijat.2011.p0115
(2011)

Paper:

Verification of Interim Check Method of CMM

Yusuke Asano, Ryoshu Furutani, and Miyu Ozaki

Tokyo Denki University, 2-2 Kanda-Nishiki-cho, Chiyoda, Tokyo 101-8457, Japan

Received:
November 29, 2010
Accepted:
December 22, 2010
Published:
March 5, 2011
Keywords:
CMM, interim check, squareness
Abstract
In industry, the Coordinate Measuring Machine (CMM) is a major geometrical measurement method because of its precision and general versatility. The CMM is conventionally used in a thermostatic chamber, but its use in the production line environment is increasing. Many factors - dust, thermal fluctuation, and collision, to name a few - deteriorate measurement precision and make it necessary to inspect the CMM daily. However, the daily inspection method is not established. Study of CMM is mainly about calibration in controlled environment like a thermostatic chamber. It is unfeasible to calibrate the CMM everyday because of its cost. So some researchers have studied about daily inspection of CMM. This paper describes daily CMM inspection procedures and how to evaluate these uncertainty. This method treats axis scale errors and interaxis squareness errors from geometric errors of CMM. These factors influence the measurand. QuickCheck (Trapet Precision Engineering) is used as an artifact. In the linear model using QuickCheck, influences of alignment error is evaluated parameters.
Cite this article as:
Y. Asano, R. Furutani, and M. Ozaki, “Verification of Interim Check Method of CMM,” Int. J. Automation Technol., Vol.5 No.2, pp. 115-119, 2011.
Data files:
References
  1. [1] T. Onishi, S. Takase, and K. Takamasu, “Study of Improvement Methods of CMM(Coordinate Measuring Machine) in Workshop Environment –Correction of Squareness Error–,” The Japan Society for Precision Engineering, Vol.73, No.7, pp. 818-822, 2007.
  2. [2] J. S. Agaplou and H. Du, “Assuring the Day-to-Day Accuracy of Coordinate Measuring Machines – A Comparison of Tools and Procedures,” J. of Manufacturing Process, Vol.9, No.2, pp. 109-120, 2007.
  3. [3] E. Curran and P. Phelan “Quick check error verification of coordinate measuring machines,” J. of Materials Processing Technology, Vol.5638, pp. 1207-1213, 2004.

*This site is desgined based on HTML5 and CSS3 for modern browsers, e.g. Chrome, Firefox, Safari, Edge, Opera.