Abstract
A low-cost and high-performance portable Doppler blood flow analysis device, which is based on a digital signal processor (DSP) TMS320V549 (Texas Instruments), contains a 240 * 320 LCD color graphic display module (Hantronix) and a thermal printer (Seiko Instruments), is developed in this study. The complex real-time autoregressive (AR) modeling is implemented in this device to estimate the time frequency representation of blood flow signals. Directional Doppler spectrograms are computed directly from the in-phase and quardrature components of the Doppler signal. Sampling frequency can vary among 5kHz, 10kHz, 15kHz, 20kHz and 25kHz to optimize the displaying dynamic range according to the blood velocity. In order to increase the display quality and provide more comprehensive information about the components of the blood flow profile, The Doppler spectrograms can be displayed in real-time on the LCD in 256 colors. They can also be printed out in 13 gray levels from the thermal printer for recording. The Doppler spectrograms computed by the AR modeling are compared with those by the STFT. The results show that this compact, economic, versatile bi-directional and battery- or line- operated Doppler device, which offers the high-performance spectral estimation and output, will be useful at different conditions, including bed-side hospitals and clinical offices.
This work was supported by the Yunnan Science and Technology Council under the Grant 2002C002.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
References
Cannon, S.R., Richards, K.L., Rollwitz, W.T.: Digital Fourier techniques in the diagnosis and quantification of aortic stenosis with pulse-Doppler echocardiography. J. Clin. Ultrasound 10, 101–107 (1982)
Kalman, P.G., Johnston, K.W., Zuech, P., Kassam, M., Poots, K.: In vitro comparison of alternative methods for quantifying the severity of Doppler spectral broadening for the diagnosis of carotid arterial occlusive disease. Ultrasound Med. Biol. 11, 435–440 (1985)
Young, J.B., Quinones, M.A., Waggoner, A.D., Miller, R.R.: Diagnosis and quantification of aortic stenosis with pulsed Doppler echocardiography. Amer. J. Cardiol. 45, 987–994 (1980)
Guo, Z., Durand, L.G., Lee, H.C.: Comparison of time-frequency distribution techniques for analysis of simulated Doppler ultrasound signals of the femoral artery. IEEE Trans. Biomed. Eng. 41, 332–342 (1994)
Kitney, R.I., Giddens, D.P.: Analysis of blood velocity waveforms by phase shift averaging and autoregressive spectral estimation. J. Biomech. Eng. 105, 398–401 (1983)
Kaluzynski, K.: Analysis of application possibilities of autoregressive modeling to Doppler blood flow signal spectral analysis. Med. Biol. Eng. Comput. 25, 373–376 (1987)
Vaitkus, P.J., Cobbold, R.S.C.: A comparative study and assessment of Doppler ultrasound spectral estimation techniques. Part I: Estimation methods. Ultrasound Med. Biol. 14, 661–672 (1988)
Vaitkus, P.J., Cobbold, R.S.C.: A comparative study and assessment of Doppler ultrasound spectral estimation techniques. Part II: methods and results. Ultrasound Med. Biol. 14, 673–688 (1988)
Wang, Y., Fish, P.J.: Comparison of Doppler signal analysis techniques for velocity waveform, turbulence and vortex measurement: a simulation study. Ultrasound Med. Biol. 22, 635–649 (1996)
Güler, N.F., Kiymik, M.K., Güler, I.: Comparison of FFT- and AR- based sonogram outputs of 20 MHz pulsed Doppler data in real time. Comput. Biol. Med. 25, 383–391 (1995)
Schlindwein, F.S., Evans, D.H.: A real-time autoregressive spectrum analyzer for Doppler ultrasound signals. Ultrasound in Med. & Biol. 15, 263–272 (1989)
Guo, Z., Durand, L.G., Allard, L., Cloutier, G., Lee, H.C., Langlois, Y.E.: Cardiac Doppler blood-flow signal analysis: Part 2 Time frequency representation based on autoregressive modeling. Med. & Biol. Eng. & Comput. 31, 242–248 (1993)
Zhang, Y., Guo, Z., Wang, W., He, S., Lee, T., Loew, M.A.: comparison of the wavelet and short-time Fourier transforms for Doppler spectral analysis. Medical Engineering and Physics 25, 547–557 (2003)
Kay, S.M., Marple, S.L.: Spectrum analysis—a modern perspective. Proc. IEEE 1969, 1380–1419 (1981)
Yu, Q., Parker, K.J.: Stochastic screen halftoning for electronic image devices. Journal of Visual Communication and Image Representation 8, 423–440 (1997)
TMS320C54x Assembly Language Tools User’s Guide, Literature Number: SPRU102C (1998)
The Staff of the Texas Instruments C5000 DSP Software Application Group, TMS320C54x DSPLIB User’s Guide (2001)
Evans, D.H., McDicken, W.N.: Doppler Ultrasound: Physics, Instrumentation and Signal Processing, 2nd edn. John Wiley & Sons, Ltd, England (2000)
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2005 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Zhang, Y., Zhou, Y., Chen, J., Shi, X., Guo, Z. (2005). A Portable Doppler Device Based on a DSP with High- Performance Spectral Estimation and Output. In: Srikanthan, T., Xue, J., Chang, CH. (eds) Advances in Computer Systems Architecture. ACSAC 2005. Lecture Notes in Computer Science, vol 3740. Springer, Berlin, Heidelberg. https://doi.org/10.1007/11572961_11
Download citation
DOI: https://doi.org/10.1007/11572961_11
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-29643-0
Online ISBN: 978-3-540-32108-8
eBook Packages: Computer ScienceComputer Science (R0)