Abstract
In the Netherlands a number of (screening) trials with digital mammography equipment have been started since 1999. In this paper results from the weekly QC procedure are given. It seems that the homogeneity test as described in the addendum to the European protocol is able to detect detector problems and flat field calibration problems. However, visual inspection remains necessary. For the CR system in the trials the homogeneity test did not find many problems. Either the homogeneity test is not effective and other tests might be more appropriate or this CR system does not have relevant image quality variations and therefore might not require weekly quality control.
Preview
Unable to display preview. Download preview PDF.
Similar content being viewed by others
References
van Engen, R., Young, K., Bosmans, H., Thijssen, M.: Addendum on digital mammography to chapter 3 of: European guidelines for Quality assurance in mammography screening, version 1.0, Euref (2002), http://www.euref.org
Oostveen, L.J., van Engen, R.E., Visser, R., Thijssen, M.A.O.: Technical quality control in digital mammography. In: Pisano, E. (ed.) Proceedings IWDM 2004, Chapel Hill (2005)
Zhao, B., Zhao, W.: Temporal performance of amorphous selenium mammography detectors. Medical Physics 32(1) (2005)
Siewerdsen, J.H., Jaffray, D.A.: Spatio-temporal response characteristics of an indirect-detection flat-panel imager. Medical Physics 26(8), 1624–1641 (1999)
Kasap, S.O., Rowlands, J.A.: Review X-ray photoconductors and stabilized a-Se for direct conversion digital flat panel X-ray image-detectors. Journal of materials science: materials in electronics (11), 179–198 (2000)
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2006 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
van Engen, R.E., Swinkels, M.M.J., Oostveen, L.J., Geertse, T.D., Visser, R. (2006). Using a Homogeneity Test as Weekly Quality Control on Digital Mammography Units. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds) Digital Mammography. IWDM 2006. Lecture Notes in Computer Science, vol 4046. Springer, Berlin, Heidelberg. https://doi.org/10.1007/11783237_36
Download citation
DOI: https://doi.org/10.1007/11783237_36
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-35625-7
Online ISBN: 978-3-540-35627-1
eBook Packages: Computer ScienceComputer Science (R0)