Abstract
To compare image quality and dose of a photon-counting multi-slit scanner (PC) and a system based on amorphous silicon (aSi), images of the CDMAM 3.4 were taken in standard mode. For 3cm PMMA, the PC used 29kV/11.7mAs, aSi used 27kV/ 50mAs. For 5cm PMMA, PC used 35kV /14.8mAs, aSi used 31kV/ 50 mAs. Exposure was manually increased for PC and lowered for aSi-system. Average glandular dose and an image quality index (IQI) were calculated over the diameter ranges 0.06 – 2.0mm and 0.1-1. In standard mode with 3cm PMMA, IQI for PC was 35% lower than for aSi at 80% lower dose. Increased dose of PC resulted in 13% lower IQI at 57% lower dose. With 5cm PMMA IQI in standard mode was 18 % lower with PC at a 69% lower dose. Increasing the dose of PC resulted a 7% lower IQI at 54% lower dose. In conclusion the PC-system might reduce dose by up to 54% at equivalent image quality, although maximal quality of aSi could not be reached.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
References
Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., Kotre, C.J.: Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys. Med. Biol. 45(11), 3225–3240 (2000)
Hemdal, B., Andersson, I., Grahn, A., Håkansson, M., Ruschin, M., Thilander-Klang, A., Båth, M., Börjesson, S., Medin, J., Tingberg, A., Månsson, L.G., Mattsson, S.: Can the average glandular dose in routine digital mammography screening be reduced? A pilot study using revised image quality criteria. Radiation Protection Dosimetry 114(1-3), 383–388 (2005)
Hermann, K.P., Obenauer, S., Funke, M., Grabbe, E.H.: Magnification mammography: a comparison of full-field digital mammography and screen-film mammography for the detection of simulated small masses and microcalcifications. Eur. Radiol. 12, 2188–2191 (2002)
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2006 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Fischmann, A., Steidle, G. (2006). Image Quality of a Photon-Counting Mammography System Compared to Digital Mammography Based on Amorphous Silicon with CsI-Scintillator. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds) Digital Mammography. IWDM 2006. Lecture Notes in Computer Science, vol 4046. Springer, Berlin, Heidelberg. https://doi.org/10.1007/11783237_59
Download citation
DOI: https://doi.org/10.1007/11783237_59
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-35625-7
Online ISBN: 978-3-540-35627-1
eBook Packages: Computer ScienceComputer Science (R0)