Image Quality of a Photon-Counting Mammography System Compared to Digital Mammography Based on Amorphous Silicon with CsI-Scintillator

  • Conference paper
Digital Mammography (IWDM 2006)

Part of the book series: Lecture Notes in Computer Science ((LNIP,volume 4046))

Included in the following conference series:

  • 1418 Accesses

Abstract

To compare image quality and dose of a photon-counting multi-slit scanner (PC) and a system based on amorphous silicon (aSi), images of the CDMAM 3.4 were taken in standard mode. For 3cm PMMA, the PC used 29kV/11.7mAs, aSi used 27kV/ 50mAs. For 5cm PMMA, PC used 35kV /14.8mAs, aSi used 31kV/ 50 mAs. Exposure was manually increased for PC and lowered for aSi-system. Average glandular dose and an image quality index (IQI) were calculated over the diameter ranges 0.06 – 2.0mm and 0.1-1. In standard mode with 3cm PMMA, IQI for PC was 35% lower than for aSi at 80% lower dose. Increased dose of PC resulted in 13% lower IQI at 57% lower dose. With 5cm PMMA IQI in standard mode was 18 % lower with PC at a 69% lower dose. Increasing the dose of PC resulted a 7% lower IQI at 54% lower dose. In conclusion the PC-system might reduce dose by up to 54% at equivalent image quality, although maximal quality of aSi could not be reached.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 84.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., Kotre, C.J.: Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys. Med. Biol. 45(11), 3225–3240 (2000)

    Article  Google Scholar 

  2. Hemdal, B., Andersson, I., Grahn, A., Håkansson, M., Ruschin, M., Thilander-Klang, A., Båth, M., Börjesson, S., Medin, J., Tingberg, A., Månsson, L.G., Mattsson, S.: Can the average glandular dose in routine digital mammography screening be reduced? A pilot study using revised image quality criteria. Radiation Protection Dosimetry 114(1-3), 383–388 (2005)

    Article  Google Scholar 

  3. Hermann, K.P., Obenauer, S., Funke, M., Grabbe, E.H.: Magnification mammography: a comparison of full-field digital mammography and screen-film mammography for the detection of simulated small masses and microcalcifications. Eur. Radiol. 12, 2188–2191 (2002)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2006 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Fischmann, A., Steidle, G. (2006). Image Quality of a Photon-Counting Mammography System Compared to Digital Mammography Based on Amorphous Silicon with CsI-Scintillator. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds) Digital Mammography. IWDM 2006. Lecture Notes in Computer Science, vol 4046. Springer, Berlin, Heidelberg. https://doi.org/10.1007/11783237_59

Download citation

  • DOI: https://doi.org/10.1007/11783237_59

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-35625-7

  • Online ISBN: 978-3-540-35627-1

  • eBook Packages: Computer ScienceComputer Science (R0)

Publish with us

Policies and ethics