Abstract
Previous studies and surgeon interviews have shown that most surgeons prefer quality standard definition (SD)TV 2D scopes to first generation 3D endoscopes. The use of a telesurgical system has eased many of the design constraints on traditional endoscopes, enabling the design of a high quality SDTV 3D endoscope and an HDTV endoscopic system with outstanding resolution. The purpose of this study was to examine surgeon performance and preference given the choice between these. The study involved two perceptual tasks and four visual-motor tasks using a telesurgical system using the 2D HDTV endoscope and the SDTV endoscope in both 2D and 3D mode. The use of a telesurgical system enabled recording of all the subjects motions for later analysis. Contrary to experience with early 3D scopes and SDTV 2D scopes, this study showed that despite the superior resolution of the HDTV system surgeons performed better with and preferred the SDTV 3D scope.
Chapter PDF
Similar content being viewed by others
Keywords
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
References
Dion, Y.M., Gaillard, F.: Visual integration of data and basic motor skills under laparoscopy. Influence of 2-D and 3-D video-camera systems. Surg. Endosc. 11, 995–1000 (1997)
Buess, G.F., van Bergen, P., Kunert, W., Schurr, M.O.: Comparative study of variuos 2D and 3D vision systems in minimally invasive surgery. Chirurg 67, 1041–1046 (1996)
Von Pichler, C., Radermacher, K., Boeckmann, W., Rau, G., Jakse, G.: Stereoscopic Visualization in Endoscopic Surgery: Problems, Benefits and potentials. Perception 6, 198–217 (1997)
Hanna, G.B., Shimi, S.M., Cuschieri, A.: Randomised study of influence of two dimensional versus three dimensional imaging on performance of laparoscopic procedures. Lancet 351, 248–251 (1998)
Jones, D.B., Brewer, J.D., Soper, N.J.: The influence of three-dimensional video systems on laparoscopic task performance. Surg. Laparosc. 6, 191–197 (1996)
Chan, A.C., Chung, S.C., Yim, A.P., Lau, J.Y., Ng, E.K., Li, A.K.: Comparison of two-dimensional vs three-dimensional camera systems in laparoscopic surgery. Surg. Endosc. 11, 438–440 (1997)
Guthart, G.S., Salisbury, J.K.: The Intuitive Telesurgery System: Overview and Application. In: Proc. of the IEEE International Conference on Robotics and Automation (ICRA 2000), San Francisco CA (April 2000)
Falk, V., Mintz, D., Reitz, B.A.: Influence of 3D-Vision on Telemanipulator Performance (2000) (submitted for publication)
Frisen, L., Lindblom, B.: Binocular summation in humans: Evidence for a hirachic model. J. Physiol. 402, 773–782 (1988)
Rabin, J.: Two eyes are better than one: binocular enhancement in the contrast domain. Opthalmic. Physiol. Opt. 15, 45–48 (1995)
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2000 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Mintz, D., Falk, V., Salisbury, J.K. (2000). Comparison of Three High-End Endoscopic Visualization Systems on Telesurgical Performance. In: Delp, S.L., DiGoia, A.M., Jaramaz, B. (eds) Medical Image Computing and Computer-Assisted Intervention – MICCAI 2000. MICCAI 2000. Lecture Notes in Computer Science, vol 1935. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-40899-4_39
Download citation
DOI: https://doi.org/10.1007/978-3-540-40899-4_39
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-41189-5
Online ISBN: 978-3-540-40899-4
eBook Packages: Springer Book Archive