Abstract
The majority of contemporary hydrocephalus valves are designed to introduce a low resistance to flow into the cerebrospinal fluid (CSF) drainage pathway, and an therefore intended to stabilise intracranial pressure (ICP) at a level close to the shunt's operating pressure. However, this goal cannot always be attained. Accelerated CSF drainage with vertical body posture in ventriculo-peritoneal shunts is one reason for the ICP decreasing below the shunt's operating pressure. Another possible factor has been studied: the impact of the pulsating pattern in the ICP on the operating pressure. Six popular constructions of medium-pressure valves were studied (Radionics Low-profile, Delta, Hakim Precision, Holter, Integra In-line and Hakim NMT). Valves were mounted in the testing rig in the UK. Shunt Evaluation Laboratory and perfused with de-ionised water at a rate of 0.3 ml min−1, and proximal pulsating pressures of different amplitudes (from 2 to 30 mmHg peak-to-peak) and frequencies (70–10 cycles min−1) were superimposed. Laboratory findings were compared with clinical material containing recordings of ICP made in patients to diagnose reasons for ventriculomegaly. The mean operating pressure decreased in all valves when the simulated amplitude of heart pulsations increased. The rate of this decrease was dependent on the type of valve (variable from 2.5 to 5 mm Hg per increase in peak-to-peak amplitude by 10 mm Hg). The decrease was not related to the frequency of the wave. The relationship between pulse amplitude and ICP in 35 patients with blocked shunts was strong (R=0.48; p<0.03; slope 0.14) and in 25 patients with properly functioning shunts was non-significant (R=0.057; p=0.765). Two examples of decrease in mean ICP in the presence of increased vasogenic ICP waves in shunted patients are presented. The shunt operating pressure, which ‘sets’ the ICP in shunted patients may be influenced by the dynamics of a patient's ICP waveform.
Similar content being viewed by others
References
Albeck, M. J., Borgesen, S. E., Gjerris, F., Schmidt, J. F., andSorensen, P. S. (1997): ‘Intracranial pressure and cerebrospinal fluid outflow conductance in healthy subjects’,J. Neurosurg.,74, pp. 597–600
Aschoff, A., Kremer, P., Benesch, C., Fruh, K., Klank, A., andKunze, S. (1995): ‘Overdrainage and shunt technology’,Child's Nerv. Syst.,11, pp. 193–202
Avezaat, C. J. J., van Eijndhoven, J. H. M., andWyper, D. J. (1979): ‘Cerebrospinal pulse-pressure and intracranial volume-pressure relationships’,J. Neurol. Neursurg. Psychiat.,42, pp. 687–700
Czosnyka, M., Wollk-Laniewski, P., Batorski, L., andZaworski, W. (1988): ‘Analysis of intracranial pressure waveform during infusion test’,Acta Neurochir (Wien),93, pp. 140–145
Czosnyka, Z., Czosnyka, M., Richards, H. K., andPickard, J. D. (1998): ‘Posture-related overdrainage: comparison of the performance of 10 hydrocephalus shuntsin vitro’,Neurosurgery,42, pp. 327–333
Czosnyka, Z. H., Czosnyka, M., andPickard, J. D. (2002): ‘Shunt testingin-vivo: a method based on the data from the UK Shunt Evaluation Laboratory’,Acta Neurochir. 81, pp. 27–30
Davson, H., Welch, K., andSegal, M. B. (1987): ‘The physiology and pathophysiology of cerebrospinal fluid’ (Churchill Livingstone, New York, 1987)
Lundkvist, B., Eklund, A., Kristensen, B., Fagerlund, M., Koskinen, L. O., andMalm, J. (2001): ‘Cerebrospinal fluid hydrodynamics after placement of a shunt with an antisiphon device: a long-term study’,J. Neurosurg,94, pp. 750–756
Marmarou, A., Shulman, K., andRosende, R. M. (1978): ‘A non-linear analysis of CSF system and intracranial pressure dynamics’,J. Neurosurg,48, pp. 332–344
Marmarou, A., Maset, A. L., Ward, J. D., Choi, S., Brooks, D., Lutz, H. A., Moulton, R. J., Muizelaar, J. P., Desalles, A., andYoung, H. F. (1987): ‘Contribution of CSF and vascular factors to elevation of ICP in severely head injured patients’,J. Neurosurg.,66, pp. 883–890
Momjian, S., Czosnyka, Z., Czosnyka, M., andPickard, J. D. (2004): ‘Link between vasogenic waves of intracranial pressure and cerebrospinal fluid outflow resistance in normal pressure hydrocephalus’,Br. J. Neurosurg.,18, pp. 56–61
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Czosnyka, Z.H., Cieslicki, K., Czosnyka, M. et al. Hydrocephalus shunts and waves of intracranial pressure. Med. Biol. Eng. Comput. 43, 71–77 (2005). https://doi.org/10.1007/BF02345125
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02345125