Abstract
This paper describes the development of a needle insertion manipulator for stereotactic neurosurgery. This robot fulfils the requirements of having both a safe mechanical design and the capacity for being sterilized. Many kinds of robots are examined in neurosurgery. Their purpose is the precise positioning of surgical instruments such as biopsy needles, electrodes etc. Some are already available commercially and have been proven useful in the operating theatre. However, their clinical application is limited by specific problems including cost, safety, positioning requirements, maintenance requirements. The main problems have been with the safety of the mechanical design and difficulties with sterilization and disinfecting pre-and post operatively. The manipulator described in this report achieves mechanical safety and has the capacity for cover-sheet-free sterilization. The manipulator has three major components: the main mechanical component (with 6 degrees of freedom), the torque transmission component, and the electric motor, which cannot be sterilized. The electrical parts are detachable. Using this mechanism, we can clearly separate the surgical area from the mechatronics components. In this paper, the basic design and the prototype development and testing are described.
Chapter PDF
Keywords
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
References
D. Glauser, H. Fankhauser, M. Epitaux, J.-L. Hefti, A. Jaccottet: Neurosurgical Robot Minerva: First Results and Current Developments, J. of Image Guided Surgery, 1:266–272, 1995.
F. Badano, et. al., The NEURO-SKILL Robot: a New Approach for Surgical Robot Development, Proc. MRCAS95, pp.318–323, 1995.
K. Masamune, et al., Robots for neurosurgery, Advanced Robotics, Vol.10, No. 4, pp.391–401, 1996.
K. Masamune, et al., Development of a CT-guided Neurosurgical Manipulator for Photon Radiosurgery System, CAR97, p.1035, 1997.
K. Masamune, et al., Development of a MRI Compatible Manipulator for Stereotactic Neurosurgery, Image Guided Surgery, 1995.
Yamauchi Y., Ohta Y., Dohi T., Kawamura H., Tanikawa T., Iseki H.: A needle insertion Manipulator for X-ray CT image guided neurosurgery, J.of LST, vol. 5–4, 814–821, 1993.
Kelly PJ, Goerss SJ, Kall BA: Evolution of contemporary instrumentation for computer-assisted stereotactic surgery, Surgical Neurology 30(3):204–15, 1988
Author information
Authors and Affiliations
Editor information
Rights and permissions
Copyright information
© 1998 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Masamune, K., Ji, L.H., Suzuki, M., Dohi, T., Iseki, H., Takakura, K. (1998). A newly developed stereotactic robot with detachable drive for neurosurgery. In: Wells, W.M., Colchester, A., Delp, S. (eds) Medical Image Computing and Computer-Assisted Intervention — MICCAI’98. MICCAI 1998. Lecture Notes in Computer Science, vol 1496. Springer, Berlin, Heidelberg. https://doi.org/10.1007/BFb0056204
Download citation
DOI: https://doi.org/10.1007/BFb0056204
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-65136-9
Online ISBN: 978-3-540-49563-5
eBook Packages: Springer Book Archive