Abstract
Purpose
Freehand tracked ultrasound imaging is an inexpensive non-invasive technique used in several guided interventions. This technique requires spatial calibration between the tracker and the ultrasound image plane. Several calibration devices (a.k.a. phantoms) use N-wires that are convenient for automatic procedures since the segmentation of fiducials in the images and the localization of the middle wires in space are straightforward and can be performed in real time. The procedures reported in literature consider only the spatial position of the middle wire. We investigate if better results can be achieved if the information of all the wires is equally taken into account. We also evaluated the precision and accuracy of the implemented methods to allow comparison with other methods.
Methods
We consider a cost function based on the in-plane errors between the intersection of all the wires with the image plane and their respective segmented points in the image. This cost function is minimized iteratively starting from a seed computed with a closed-form solution based on the middle wires.
Results
Mean calibration precision achieved with the N-wire phantom was about 0.5 mm using a shallow probe, and mean accuracy was around 1.4 mm with all implemented methods. Precision was about 2.0 mm using a deep probe.
Conclusions
Precision and accuracy achieved with the N-wire phantom and a shallow probe are at least comparable to that obtained with other methods traditionally considered more precise. Calibration using N-wires can be done more consistently if the parameters are optimized with the proposed cost function.







Similar content being viewed by others
References
Brown RA et al (1979) A stereotactic head frame for use with CT body scanners. Inves Radiol 14(4):300
Chatrasingh M, Suthakorn J, Ongwattanakul S, Wiratkapun C, Shuijantuq S Ultrasound calibration toolkit with a high-adjustability feature based on user requirements
Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med biol 35(1):79–93
Comeau RM, Fenster A, Peters TM (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Medical Imaging’98, pp 747–754. International Society for Optics and Photonics, San Francico
Laboratory for Percutaneous Surgery (PerkLab) at Queen’s University. Plus (public software library for ultrasound imaging research). Internet development repository, http://plustoolkit.org. Accessed January (2013)
Hsu P-W, Prager RW, Gee AH, Treece GM (2007) Freehand 3d ultrasound calibration: A review. Technical report. University of Cambridge, Cambridge
Lasso A, Heffter T, Pinter CS, Ungi T, Fichtinger G (2012) Implementation of the PLUS open-source toolkit for translational research of ultrasound-guided intervention systems. The MIDAS Journal—Systems and Architectures for Computer Assisted Interventions 2012, Workshop at Medical Image Computing and Computer-Assisted Intervention (MICCAI 2012), pp 1–12, http://hdl.handle.net/10380/3367
Lee S, Fichtinger G, Chirikjian GS (2002) Numerical algorithms for spatial registration of line fiducials from cross-sectional images. Med Phys 29:1881
Lindseth F, Tangen GA, Lango T, Bang J et al (2003) Probe calibration for freehand 3D ultrasound. Ultrasound Med Biol 29(11):1607
Mercier L, Lang T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3D ultrasound systems. Ultrasound Med Biol 31(2):143–165
Muratore DM, Galloway RL (2001) Beam calibration without a phantom for creating a 3D freehand ultrasound system. Ultrasound Med Biol 27(11):1557–1566
Pagoulatos N, Haynor DR, Kim Y (1999) Fast calibration for 3D ultrasound imaging and multimodality image registration. In: Engineering in Medicine and Biology, 1999. 21st Annual Conf. and the 1999 Annual Fall Meeting of the Biomedical Engineering Soc., BMES/EMBS Conference, 1999. Proceedings of the First Joint, vol 2, p 1065
Prager RW, Rohling RN, Gee AH, Berman L et al (1998) Rapid calibration for 3D freehand ultrasound. Ultrasound Med Biol 24(6):855
Pieper S, Halle M, Kikinis R (2004) 3D SLICER. In: Proceedings of the 1st IEEE International Symposium on Biomedical Imaging: From Nano to Macro, pp 632–635
Acknowledgments
This work was supported in part by Agencia Nacional de Investigacion e Innovacion (ANII, Uruguay) under grant (BE POS 2010 2236). Comision Sectorial de Investigacion Cientıfica (CSIC, Universidad de la Republica, Uruguay) supported the internship of G. Carbajal at Perklab. This work was co-funded as an Applied Cancer Research Unit of Cancer Care Ontario with funds provided by the Ontario Ministry of Health and Long-Term Care. G. Fichtinger is supported as a Cancer Care Ontario Research Chair.
Conflict of interest None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Carbajal, G., Lasso, A., Gómez, Á. et al. Improving N-wire phantom-based freehand ultrasound calibration. Int J CARS 8, 1063–1072 (2013). https://doi.org/10.1007/s11548-013-0904-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-013-0904-9