Target visibility enhancement for C-arm cone beam CT-fluoroscopy-guided hepatic needle placement: implementation and accuracy evaluation

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

Fluoroscopy-guided hepatic intervention is limited by target visibility and respiratory movement. A feasible procedure for visibility enhancement of the key regions and targets in 2D fluoroscopic images is needed. A system was developed to improve targeting by integrating the forward projection of objects extracted from 3D cone beam CT (CBCT) volumes. The target matching accuracy during regular respiration was measured to evaluate the system.

Method

3D CBCT abdominal volumes were acquired and segmented to extract different regions, including the diaphragm, hepatic vessels, bony structures, and hepatic tumor. The segmented result was rendered and projected to generate augmented fluoroscopy fusion images. The target matching accuracy by applying these procedures was evaluated for the hepatic intervention guidance.

Result

Quantitative assessment of the target matching accuracy in the upper section of liver was performed for eight targets from four subjects. The 2D and 3D target matching accuracy were \(0.98\,\pm \,0.37\) and \(1.47\,\pm \,0.26\) mm, respectively. The 2D target matching accuracy was \(1.46\,\pm \,0.67\) mm for the target in the lower liver. This accuracy should be acceptable for the 5 mm safety margin required in clinical use.

Conclusion

Visibility of targets in 2D fluoroscopy was enhanced to improve interactive navigation guidance for hepatic needle placement. The target matching accuracy for the C-arm cone beam CT-fluoroscopy-guided hepatic needle targeting was sufficient for clinical use.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Siewerdsen JH, Moseley DJ, Burch S, Bisland SK, Bogaards A, Wilson BC, Jaffray DA (2005) Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. Med Phys 32:241

    Article  CAS  PubMed  Google Scholar 

  2. Braak SJ, Van Strijen MJL, Van Leersum M, Van Es HW, Van Heesewijk JPM (2010) Real-time 3D fluoroscopy guidance during needle interventions: technique, accuracy, and feasibility. Am J Roentgenol 194(5):W445–W451

    Article  CAS  Google Scholar 

  3. Braak SJ, Herder GJ, van Heesewijk JP, van Strijen MJ (2012) Pulmonary masses: Initial results of cone-beam CT guidance with needle planning software for percutaneous lung biopsy. Cardiovasc Intervent Radiol 35(6):1414–1421

    Article  PubMed  Google Scholar 

  4. Jin KN, Park CM, Goo JM, Lee HJ, Lee Y et al (2010) Initial experience of percutaneous transthoracic needle biopsy of lung nodules using C-arm cone-beam CT systems. Eur Radiol 20(9):2108–2115

    Article  PubMed  Google Scholar 

  5. Hirota S, Nakao N, Yamamoto S, Kobayashi K, Maeda H et al (2006) Cone-beam CT with flat-panel-detector digital angiography system: early experience in abdominal interventional procedures. Cardiovasc Intervent Radiol 29(6):1034–1038

    Article  PubMed  Google Scholar 

  6. Morimoto M, Numata K, Kondo M, Nozaki A, Hamaguchi S, Takebayashi S, Tanaka K (2010) C-arm cone beam CT for hepatic tumor ablation under real-time 3D imaging. Am J Roentgenol 194(5):W452–W454

    Article  Google Scholar 

  7. Kroeze SG, Huisman M, Verkooijen HM, van Diest PJ, Bosch JR, van den Bosch MA (2012) Real-time 3D fluoroscopy-guided large core needle biopsy of renal masses: a critical early evaluation according to the IDEAL recommendations. Cardiovasc Intervent Radiol 35(3):680–685

    Article  PubMed  Google Scholar 

  8. Pedicelli A, Verdolotti T, Pompucci A, Desiderio F et al (2011) Interventional spinal procedures guided and controlled by a 3D rotational angiographic unit. Skeletal Radiol 40(12):1595–1601

    Article  PubMed  Google Scholar 

  9. Powell MF, DiNobile D, Reddy AS (2010) C-arm fluoroscopic cone beam CT for guidance of minimally invasive spine interventions. Pain Physician 13(1):51–59

    PubMed  Google Scholar 

  10. Leschka SC, Babic D, El Shikh S, Wossmann C, Schumacher M, Taschner CA (2012) C-arm cone beam computed tomography needle path overlay for image-guided procedures of the spine and pelvis. Neuroradiology 54(3):215–223

  11. Racadio JM, Babic D, Homan R, Rampton JW, Patel MN, Racadio JM, Johnson ND (2007) Live 3D guidance in the interventional radiology suite. Am J Roentgenol 189(6):W357–W364

  12. Schumann C, Bieberstein J, Braunewell S, Niethammer M, Peitgen HO (2012) Visualization support for the planning of hepatic needle placement. Int J Comput Assist Radiol Surg 7(2):191–197

  13. Baegert C, Villard C, Schreck P, Soler L (2007) Multi-criteria trajectory planning for hepatic radiofrequency ablation. In Med Image Comput Comput Assist Interv-MICCAI 2007:676–684

    Google Scholar 

  14. Kass M, Witkin A, Terzopoulos D (1988) Snakes: active contour models. Int J Comput Vision 1(4):321–331

    Article  Google Scholar 

  15. Xu C, Prince JL (1998) Snakes, shapes, and gradient vector flow. IEEE Trans Image Process 7(3):359–369

    Article  CAS  PubMed  Google Scholar 

  16. Sethian JA (1999) Level set methods and fast marching methods: evolving interfaces in computational geometry, fluid mechanics, computer vision, and materials science. Cambridge University Press, Cambridge

    Google Scholar 

  17. Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G (2006) User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage 31(3):1116–1128

    Article  PubMed  Google Scholar 

  18. Caselles V, Kimmel R, Sapiro G (1997) Geodesic active contours. Int J Comput Vision 22(1):61–79

    Article  Google Scholar 

  19. Frangi AF, Niessen WJ, Vincken KL, Viergever MA (1998) Multiscale vessel enhancement filtering. Med Image Comput Comput-Assist Interv-MICCAI 1998:130–137

    Google Scholar 

  20. Wang M, Ding H, Wang G (2012) An improved FDK algorithm using camera calibration technique for reconstruction of misaligned CBCT system. In Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE, pp 5991–5994

  21. Feuerstein M, Mussack T, Heining SM, Navab N (2008) Intraoperative laparoscope augmentation for port placement and resection planning in minimally invasive liver resection. IEEE Trans Med Imaging 27(3):355–369

    Article  PubMed  Google Scholar 

  22. Reaungamornrat S, Otake Y, Uneri A, Schafer S, Mirota DJ, Nithiananthan S, Stayman JW, Kleinszig G, Khanna AJ, Taylor RH, Siewerdsen JH (2012) An on-board surgical tracking and video augmentation system for C-arm image guidance. Int J Comput Assist Radiol Surg 7(5):647–665

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

This work was supported in part by grants from National Basic Research Program of China (2011CB707701), National Natural Science Foundation of China (81271671, 81127003, 51361130032).

Conflict of interest

None

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Guangzhi Wang.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, M., Ding, H., Wang, X. et al. Target visibility enhancement for C-arm cone beam CT-fluoroscopy-guided hepatic needle placement: implementation and accuracy evaluation. Int J CARS 10, 263–273 (2015). https://doi.org/10.1007/s11548-014-1070-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-014-1070-4

Keywords