Abstract
Purpose
Fluoroscopy-guided hepatic intervention is limited by target visibility and respiratory movement. A feasible procedure for visibility enhancement of the key regions and targets in 2D fluoroscopic images is needed. A system was developed to improve targeting by integrating the forward projection of objects extracted from 3D cone beam CT (CBCT) volumes. The target matching accuracy during regular respiration was measured to evaluate the system.
Method
3D CBCT abdominal volumes were acquired and segmented to extract different regions, including the diaphragm, hepatic vessels, bony structures, and hepatic tumor. The segmented result was rendered and projected to generate augmented fluoroscopy fusion images. The target matching accuracy by applying these procedures was evaluated for the hepatic intervention guidance.
Result
Quantitative assessment of the target matching accuracy in the upper section of liver was performed for eight targets from four subjects. The 2D and 3D target matching accuracy were \(0.98\,\pm \,0.37\) and \(1.47\,\pm \,0.26\) mm, respectively. The 2D target matching accuracy was \(1.46\,\pm \,0.67\) mm for the target in the lower liver. This accuracy should be acceptable for the 5 mm safety margin required in clinical use.
Conclusion
Visibility of targets in 2D fluoroscopy was enhanced to improve interactive navigation guidance for hepatic needle placement. The target matching accuracy for the C-arm cone beam CT-fluoroscopy-guided hepatic needle targeting was sufficient for clinical use.







Similar content being viewed by others
References
Siewerdsen JH, Moseley DJ, Burch S, Bisland SK, Bogaards A, Wilson BC, Jaffray DA (2005) Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. Med Phys 32:241
Braak SJ, Van Strijen MJL, Van Leersum M, Van Es HW, Van Heesewijk JPM (2010) Real-time 3D fluoroscopy guidance during needle interventions: technique, accuracy, and feasibility. Am J Roentgenol 194(5):W445–W451
Braak SJ, Herder GJ, van Heesewijk JP, van Strijen MJ (2012) Pulmonary masses: Initial results of cone-beam CT guidance with needle planning software for percutaneous lung biopsy. Cardiovasc Intervent Radiol 35(6):1414–1421
Jin KN, Park CM, Goo JM, Lee HJ, Lee Y et al (2010) Initial experience of percutaneous transthoracic needle biopsy of lung nodules using C-arm cone-beam CT systems. Eur Radiol 20(9):2108–2115
Hirota S, Nakao N, Yamamoto S, Kobayashi K, Maeda H et al (2006) Cone-beam CT with flat-panel-detector digital angiography system: early experience in abdominal interventional procedures. Cardiovasc Intervent Radiol 29(6):1034–1038
Morimoto M, Numata K, Kondo M, Nozaki A, Hamaguchi S, Takebayashi S, Tanaka K (2010) C-arm cone beam CT for hepatic tumor ablation under real-time 3D imaging. Am J Roentgenol 194(5):W452–W454
Kroeze SG, Huisman M, Verkooijen HM, van Diest PJ, Bosch JR, van den Bosch MA (2012) Real-time 3D fluoroscopy-guided large core needle biopsy of renal masses: a critical early evaluation according to the IDEAL recommendations. Cardiovasc Intervent Radiol 35(3):680–685
Pedicelli A, Verdolotti T, Pompucci A, Desiderio F et al (2011) Interventional spinal procedures guided and controlled by a 3D rotational angiographic unit. Skeletal Radiol 40(12):1595–1601
Powell MF, DiNobile D, Reddy AS (2010) C-arm fluoroscopic cone beam CT for guidance of minimally invasive spine interventions. Pain Physician 13(1):51–59
Leschka SC, Babic D, El Shikh S, Wossmann C, Schumacher M, Taschner CA (2012) C-arm cone beam computed tomography needle path overlay for image-guided procedures of the spine and pelvis. Neuroradiology 54(3):215–223
Racadio JM, Babic D, Homan R, Rampton JW, Patel MN, Racadio JM, Johnson ND (2007) Live 3D guidance in the interventional radiology suite. Am J Roentgenol 189(6):W357–W364
Schumann C, Bieberstein J, Braunewell S, Niethammer M, Peitgen HO (2012) Visualization support for the planning of hepatic needle placement. Int J Comput Assist Radiol Surg 7(2):191–197
Baegert C, Villard C, Schreck P, Soler L (2007) Multi-criteria trajectory planning for hepatic radiofrequency ablation. In Med Image Comput Comput Assist Interv-MICCAI 2007:676–684
Kass M, Witkin A, Terzopoulos D (1988) Snakes: active contour models. Int J Comput Vision 1(4):321–331
Xu C, Prince JL (1998) Snakes, shapes, and gradient vector flow. IEEE Trans Image Process 7(3):359–369
Sethian JA (1999) Level set methods and fast marching methods: evolving interfaces in computational geometry, fluid mechanics, computer vision, and materials science. Cambridge University Press, Cambridge
Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G (2006) User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage 31(3):1116–1128
Caselles V, Kimmel R, Sapiro G (1997) Geodesic active contours. Int J Comput Vision 22(1):61–79
Frangi AF, Niessen WJ, Vincken KL, Viergever MA (1998) Multiscale vessel enhancement filtering. Med Image Comput Comput-Assist Interv-MICCAI 1998:130–137
Wang M, Ding H, Wang G (2012) An improved FDK algorithm using camera calibration technique for reconstruction of misaligned CBCT system. In Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE, pp 5991–5994
Feuerstein M, Mussack T, Heining SM, Navab N (2008) Intraoperative laparoscope augmentation for port placement and resection planning in minimally invasive liver resection. IEEE Trans Med Imaging 27(3):355–369
Reaungamornrat S, Otake Y, Uneri A, Schafer S, Mirota DJ, Nithiananthan S, Stayman JW, Kleinszig G, Khanna AJ, Taylor RH, Siewerdsen JH (2012) An on-board surgical tracking and video augmentation system for C-arm image guidance. Int J Comput Assist Radiol Surg 7(5):647–665
Acknowledgments
This work was supported in part by grants from National Basic Research Program of China (2011CB707701), National Natural Science Foundation of China (81271671, 81127003, 51361130032).
Conflict of interest
None
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Wang, M., Ding, H., Wang, X. et al. Target visibility enhancement for C-arm cone beam CT-fluoroscopy-guided hepatic needle placement: implementation and accuracy evaluation. Int J CARS 10, 263–273 (2015). https://doi.org/10.1007/s11548-014-1070-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-014-1070-4