Robot-assisted automatic ultrasound calibration

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

Ultrasound (US) calibration is the process of determining the unknown transformation from a coordinate frame such as the robot’s tooltip to the US image frame and is a necessary task for any robotic or tracked US system. US calibration requires submillimeter-range accuracy for most applications, but it is a time-consuming and repetitive task. We provide a new framework for automatic US calibration with robot assistance and without the need for temporal calibration.

Method

US calibration based on active echo (AE) phantom was previously proposed, and its superiority over conventional cross-wire phantom-based calibration was shown. In this work, we use AE to guide the robotic arm motion through the process of data collection; we combine the capability of the AE point to localize itself in the frame of the US image with the automatic motion of the robotic arm to provide a framework for calibrating the arm to the US image automatically.

Results

We demonstrated the efficacy of the automated method compared to the manual method through experiments. To highlight the necessity of frequent ultrasound calibration, it is demonstrated that the calibration precision changed from 1.67 to 3.20 mm if the data collection is not repeated after a dismounting/mounting of the probe holder. In a large data set experiment, similar reconstruction precision of automatic and manual data collection was observed, while the time was reduced by 58 %. In addition, we compared ten automatic calibrations with ten manual ones, each performed in 15 min, and showed that all the automatic ones could converge in the case of setting the initial matrix as identity, while this was not achieved by manual data sets. Given the same initial matrix, the repeatability of the automatic was [0.46, 0.34, 0.80, 0.47] versus [0.42, 0.51, 0.98, 1.15] mm in the manual case for the US image four corners.

Conclusions

The submillimeter accuracy requirement of US calibration makes frequent data collections unavoidable. We proposed an automated calibration setup and showed feasibility by implementing it for a robot tooltip to US image calibration. The automated method showed a similar reconstruction precision as well as repeatability compared to the manual method, while the time consumed for data collection was reduced. The automatic method also reduces the burden of data collection for the user. Thus, the automated method can be a viable solution for applications that require frequent calibrations.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Ungi T, Lasso A, Fichtinger G (2015) Tracked US in navigated spine interventions. In: Li S, Yao J (eds) Spinal imaging and image analysis. Springer International Publishing, pp 469–494

  2. Westwood JD (2014) An US-based navigation system for minimally invasive neck surgery. Med Meets Virtual Real 21 NextMed/MMVR21 196:36

    Google Scholar 

  3. Monfaredi R, Wilson E, Azizi koutenaei B, Labrecque B, Leroy K, Goldie J, Louis E, Swerdlow D, Cleary K (2014) Robot-assisted US imaging: overview and development of a parallel telerobotic system. Minim Invasive Ther Allied Technol 1–9

  4. Sauer F, Khamene A, Bascle B, Schinunang L, Wenzel F, Vogt S (2001) Augmented reality visualization of ultrasound images: system description, calibration, and features. In Augmented reality, 2001. Proceedings IEEE and ACM international symposium on, pp 30–39. IEEE

  5. Mebarki R, Krupa A, Chaumette F (2010) 2-d US probe complete guidance by visual servoing using image moments. Robot IEEE Trans 26(2):296–306

    Article  Google Scholar 

  6. Chen TK, Leung C, Azar RZ, Chan K-K, Zhuang B, Dickie K, Dixon J, Pendziwol L, Pelissier L (2012) Laurent importance of transducer position tracking for automated breast US: initial assessments. In: Ultrasonics symposium (IUS), 2012 IEEE International. IEEE

  7. Aalamifar F, Jiang D, Zhang HK, Cheng A, Guo X, Khurana R, Iordachita I, Boctor EM (2015) Co-robotic US tomography: dual arm setup and error analysis. In: Bosch JG, Duric N (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 94190N

  8. Mercier L, Langø T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3-D US systems. US Med biol 31(2):143–165

    Article  Google Scholar 

  9. Boctor EM, Iordachita I, Choti MA, Hager G, Fichtinger G (2006) Bootstrapped ultrasound calibration. Stud Health Technol Inform 119:61–66

    PubMed  Google Scholar 

  10. Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35(1):79–93

    Article  PubMed  Google Scholar 

  11. Chen TK, Thurston AD, Moghari MH, Ellis RE, Abolmaesumi P (2008) A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness. In: Miga MI, Cleary KR (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 69182A

  12. Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: Biomedical imaging: nano to macro, 2004. IEEE international symposium on, pp. 527–530. IEEE

  13. Viswanathan A, Boctor EM, Taylor RH, Hager G, Fichtinger G (2004) Immediate ultrasound calibration with three poses and minimal image processing. In: Barillot C, Haynor DR, Hellier P (eds) Medical image computing and computer-assisted intervention-MICCAI 2004. Springer, Berlin, pp 446–454

  14. Dandekar S, Li Y, Molloy J, Hossack J (2005) A phantom with reduced complexity for spatial 3-D ultrasound calibration. Ultrasound Med Biol 31(8):1083–1093

    Article  PubMed  Google Scholar 

  15. Cheng A, Ackerman MK, Chirikjian GS, Boctor EM (2014) Design and development of an US calibration phantom and system. In: Yaniv ZR, Holmes DR (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 903624A

  16. Hsu P-W, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D US calibration. US Med Biol 34(2):239–251

    Article  Google Scholar 

  17. Carbajal G, Lasso A, Gómez Á, Fichtinger G (2013) Improving N-wire phantom-based freehand US calibration. Int J Comput Assist Radiol Surg 8(6):1063–1072

    Article  PubMed  Google Scholar 

  18. Hsu P-W, Treece GM, Prager RW, Houghton NE, Gee Andrew H (2008) Comparison of freehand 3-D US calibration techniques using a stylus. US Med Biol 34(10):1610–1621

  19. Melvær EL, Mørken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7(4):611–620

    Article  PubMed  Google Scholar 

  20. Zhang H, Banovac F, White A, Cleary K (2006) Freehand 3D ultrasound calibration using an electromagnetically tracked needle. In: Cleary KR, Galloway RL Jr (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 61412M

  21. Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3-D freehand US. US Med Biol 24(6):855–869

    Article  CAS  Google Scholar 

  22. Guo X, Cheng A, Zhang HK, Kang H-J, Etienne-Cummings R,Boctor EM (2014) Active echo: a new paradigm for US calibration. In: Golland P, Hata N, Barillot C, Hornegger J, Howe R (eds) Medical image computing and computer assisted intervention-MICCAI 2014. Springer International Publishing, pp 397–404

  23. Guo X, Kang H-J, Etienne-Cummings R, Boctor EM (2014) Active US Pattern Injection System (AUSPIS) for interventional tool guidance. PLoS One 9(10):e104262

    Article  PubMed  PubMed Central  Google Scholar 

  24. Ackerman MK, Cheng A, Boctor E, Chirikjian G (2014) Online US sensor calibration using gradient descent on the Euclidean group. In: Robotics and automation (ICRA), 2014 IEEE international conference on. IEEE

Download references

Acknowledgments

We would like to thank Dr. Lei Chen for his valuable feedback on the accuracy of calibrated probe in various experiments and Dr. Xiaoyu Guo for providing the AE circuitry. Financial support was provided by Johns Hopkins University internal funds and NSF Grant IIS-1162095.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Fereshteh Aalamifar.

Ethics declarations

Conflict of interest

Authors have no conflict of interest with the content of this manuscript.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Aalamifar, F., Cheng, A., Kim, Y. et al. Robot-assisted automatic ultrasound calibration. Int J CARS 11, 1821–1829 (2016). https://doi.org/10.1007/s11548-015-1341-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-015-1341-8