Cardiac X-ray image-based haptic guidance for robot-assisted coronary intervention: a feasibility study

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

Effective and efficient haptic guidance is desirable for tele-operated robotic surgery because it has a potential to enhance surgeon’s skills, especially in coronary interventions where surgeon loses both an eye–hand coordination and a direct sight to the organ. This paper proposes a novel haptic guidance procedure—both kinesthetic and cutaneous, which solely depends upon X-ray images, for tele-robotic system that assists an efficient navigation of the guidewire towards the target location during a coronary intervention.

Methods

Proposed methodology requires cardiologists to draw virtual fixtures (VFs) on angiograms as a preoperative procedure. During an operation, these VFs direct the guidewire to the desired coronary vessel. For this, the position and orientation of guidewire tip are calculated with respect to VFs’ anatomy, using image processing on the real-time 2D fluoroscopic images. The haptic feedbacks are then rendered on to the master device depending on the interaction with attractive and repulsive, guidance and forbidden region VFs.

Results

A feasibility study in the laboratory environment is performed by using a webcam as an image acquisition device and a phantom-based coronary vessel model. The subsequent statistical analysis shows that, on an average, a decrease of approx. 37% in task completion time is observed with haptic feedback. Moreover, haptic guidance is found effective for most difficult branch, whereas there is a minimal significance of such haptics for the easiest branch.

Conclusions

The proposed haptic guidance procedure may assist cardiologists for an efficient and effective guidewire navigation during a surgical procedure. The cutaneous haptics (vibration feedback) is found more helpful in coronary interventions compared with kinesthetic haptics (force feedback).

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig.4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Mozaffarian D, Benjamin EJ, Go AS, Arnett DK, Blaha MJ, Cushman M, Mary De Ferranti, Després JP, Fullerton HJ, Howard VJ, Huffman MD, Mark SE, Judd D, Kissela BM, Lackland DT, Lichtman JH, Lisabeth LD, Liu S, Mackey RH, Matchar DB, McGuire DK, Mohler ER, Moy CS, Muntner P, Mussolino ME, Nasir K, Neumar RW, Nichol RW, Palaniappan L, Pandey DK, Reeves MJ, Rodriguez CJ, Sorlie PD, Stein J, Towfighi A, Turan TN, Virani SS, Willey JZ, Woo JZ, Yeh RW, Turner MB (2015) Heart disease and stroke statistics-2015 update : a report from the American Heart Association, vol. 131, no. 4

  2. Baim DSWG (2001) Grossmanʼs cardiac catheterization, angiography, and intervention, 6Th Edition, vol. 15, no. 6

  3. Leaney B (2006) What’s new in vascular interventional radiology? Aortic stent grafting. Aust Fam Phys 35(5):294–297

    Google Scholar 

  4. George JC, Tabaza L, Janzer S (2020) Robotic-assisted balloon angioplasty and stent placement with distal embolic protection device for severe carotid artery stenosis in a high-risk surgical patient. Catheter Cardiovasc Interv 96(2):410–412. https://doi.org/10.1002/ccd.28939

    Article  PubMed  Google Scholar 

  5. Moon Y, Choi J (2014) A tele-operated robotic platform for cardiovascular catheter intervention with haptic guidance. Int Conf Control Autom Syst, no Iccas. https://doi.org/10.1109/ICCAS.2014.6988007

    Article  Google Scholar 

  6. Lu Q, Shen Y, Xia S, Chen B, Wang K (2020) A novel universal endovascular robot for peripheral arterial stent-assisted angioplasty: initial experimental results. Vasc Endovascular Surg 54(7):598–604. https://doi.org/10.1177/1538574420940832

    Article  PubMed  Google Scholar 

  7. Enayati N, De Momi E, Ferrigno G (2016) Haptics in robot-assisted surgery: challenges and benefits. IEEE Rev Biomed Eng 9:49–65. https://doi.org/10.1109/RBME.2016.2538080

    Article  PubMed  Google Scholar 

  8. Talvas A, Marchal M, Lecuyer A (2014) A survey on bimanual haptic interaction. IEEE Trans Haptics 7(3):285–300. https://doi.org/10.1109/TOH.2014.2314456

    Article  PubMed  Google Scholar 

  9. King CH, Culjat MO, Franco ML, Lewis CE, Dutson EP, Grundfest WS, Bisley JW (2009) Tactile feedback induces reduced grasping force in robot-assisted surgery. IEEE Trans Haptics 2(2):103–110. https://doi.org/10.1109/TOH.2009.4

    Article  PubMed  Google Scholar 

  10. “HansenMedical.” http://www.hansenmedical.com/.

  11. “Mentice.” http://www.mentice.com/.

  12. Yin X, Guo S, Hirata H, Ishihara H (2014) Design and experimental evaluation of a teleoperated haptic robot – assisted catheter operating system. J Intell Mater Syst Struct. https://doi.org/10.1177/1045389X14556167

    Article  Google Scholar 

  13. Baumann R, Clavel R (1998) Haptic interface for virtual reality based minimally invasive surgery simulation. Proc 1998 IEEE Int Conf Robot Autom Cat No98CH36146. https://doi.org/10.1109/ROBOT.1998.676435

    Article  Google Scholar 

  14. Park JW, Choi J, Pak HN, Lee SJ, Park Y, Shin SM, Sun K (2010) Development of a force-reflecting robotic platform for cardiac catheter navigation. Artif Organs 34(11):1034–1039. https://doi.org/10.1111/j.1525-1594.2010.01142.x

    Article  PubMed  Google Scholar 

  15. Xiao N, Guo P, Guo S (2015) Push force feedback for a kind of robotic catheter navigation system. 2015 IEEE Int Conf Inf Autom ICIA 2015-conjunction IEEE Int Conf Atom Logist. https://doi.org/10.1109/ICInfA.2015.7279254

    Article  Google Scholar 

  16. Kuebler B, Seibold U, Hirzinger G (2005) Development of actuated and sensor integrated forceps for minimally invasive robotic surger. Int J Med Robot 1(3):96–107. https://doi.org/10.1002/rcs.33

    Article  CAS  PubMed  Google Scholar 

  17. Meiß T, Budelmann C, Kern TA, Sindlinger S, Minamisava C, Werthschützky R (2009) Intravascular palpation and haptic feedback during angioplasty. Proc - 3rd Jt EuroHaptics Conf Symp Haptic Interfaces Virtual Environ Teleoperator Syst World Haptics https://doi.org/10.1109/WHC.2009.4810904

  18. Ganji Y, Janabi-Sharifi F, Cheema AN (2009) Robot-assisted catheter manipulation for intracardiac navigation. Int J Comput Assist Radiol Surg 4(4):307–315. https://doi.org/10.1007/s11548-009-0296-z

    Article  PubMed  Google Scholar 

  19. Hoshiar AK, Jeon S, Kim K, Lee S, Kim JY, Choi H (2018) Steering algorithm for a flexible microrobot to enhance guidewire control in a coronary angioplasty application. Micromachines 9(12):1–13. https://doi.org/10.3390/mi9120617

    Article  Google Scholar 

  20. Russo AD, Fassini G, Conti S, Casella M, Russo E, Riva S, Moltrasio M, Tundo F, Martino GD, Gallinghouse GJ (2016) Analysis of catheter contact force during atrial fibrillation ablation using the robotic navigation system : results from a randomized study. J Int Cardiac Electrophysiol. https://doi.org/10.1007/s10840-016-0102-0

    Article  Google Scholar 

  21. Park S, Howe RD, Torchiana DF (2001) Virtual fixtures for robotic cardiac surgery. Image Comput Med. https://doi.org/10.1007/3-540-45468-3_252

    Article  Google Scholar 

  22. Howard T, Szewczyk J (2014) Visuo-haptic feedback for 1-D Guidance in laparoscopic surgery. 2014 5th IEEE RAS EMBS Int Conf Biomed Robot Biomechatronics. https://doi.org/10.1109/BIOROB.2014.6913752

    Article  Google Scholar 

  23. Ryden F, Chizeck HJ (2012) Forbidden-region virtual fixtures from streaming point clouds: Remotely touching and protecting a beating heart. IEEE Int Conf Intell Robot Syst. https://doi.org/10.1109/IROS.2012.6386012

    Article  Google Scholar 

  24. Ren J, Patel RV, McIsaac KA, Guiraudon G, Peters TM (2008) Dynamic 3-D virtual fixtures for minimally invasive beating heart procedures. IEEE Trans Med Imaging 27(8):1061–1070. https://doi.org/10.1109/TMI.2008.917246

    Article  PubMed  Google Scholar 

  25. Abbott J, Panadda M, Allison O (2007) Haptic virtual fixtures for robot-assisted manipulation. Res Robot. https://doi.org/10.1007/978-3-540-48113-3_5

    Article  Google Scholar 

  26. Ji C, Hou ZG, Xie XL (2011) An image-based guidewire navigation method for robot-assisted intravascular interventions. Annu Int Conf IEEE Eng Med Biol Soc EMBS Proc. https://doi.org/10.1109/IEMBS.2011.6091647

    Article  Google Scholar 

  27. Dagnino G, Liu J, Abdelaziz MEMK, Chi W, Riga C, Yang G (2018) Haptic feedback and dynamic active constraints for robot - assisted endovascular catheterization. pp. 1770–1775

  28. Park JW, Choi J, Park Y, Sun K (2011) Haptic virtual fixture for robotic cardiac catheter navigation. Artif Organs 35(11):1127–1131. https://doi.org/10.1111/j.1525-1594.2011.01373.x

    Article  PubMed  Google Scholar 

  29. Woo J, Song H, Cha H, Yi B (2019) Applied sciences advantage of steerable catheter and haptic feedback for a 5-DOF vascular intervention robot system

  30. Lo N, Gutierrez JA, Swaminathan RV (2018) Robotic-assisted percutaneous coronary intervention. Treat Opt Cardiovasc Med Curr. https://doi.org/10.1007/s11936-018-0608-0

    Article  Google Scholar 

  31. Kim D, Park S, Jeong MH, Ryu J (2018) Registration of angiographic image on real-time fluoroscopic image for image-guided percutaneous coronary intervention. Int J Comput Assist Radiol Surg 13(2):203–213. https://doi.org/10.1007/s11548-017-1689-z

    Article  PubMed  Google Scholar 

  32. Frangi AF, Niessen WJ, Vincken KL, a Viergever M (1998) Multiscale vessel enhancement filtering. Medial Image Comput Comput Invervention – MICCAI Lect Notes Comput Sci https://doi.org/10.1016/j.media.2004.08.001

  33. Suzuki S, be KA (1985) Topological structural analysis of digitized binary images by border following. Comput Vision, Graph Image Process. https://doi.org/10.1016/0734-189X(85)90016-7

    Article  Google Scholar 

  34. Gonzalez RC, Woods RE, Hall P, Digital Image Processing

Download references

Acknowledgements

Authors acknowledge the equipment and lab resources are provided by University of Engineering and Technology, Lahore (Faisalabad Campus).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Asim Ghaffar.

Ethics declarations

Conflict of interest

Authors declare that there is no conflict of interests.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tahir, A., Iqbal, H., Usman, M. et al. Cardiac X-ray image-based haptic guidance for robot-assisted coronary intervention: a feasibility study. Int J CARS 17, 531–539 (2022). https://doi.org/10.1007/s11548-022-02563-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-022-02563-w

Keywords

Navigation