Online model checking for monitoring surrogate-based respiratory motion tracking in radiation therapy

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Objective

Correlation between internal and external motion is critical for respiratory motion compensation in radiosurgery. Artifacts like coughing, sneezing or yawning or changes in the breathing pattern can lead to misalignment between beam and tumor and need to be detected to interrupt the treatment. We propose online model checking (OMC), a model-based verification approach from the field of formal methods, to verify that the breathing motion is regular and the correlation holds. We demonstrate that OMC may be more suitable for artifact detection than the prediction error.

Materials and methods

We established a sinusoidal model to apply OMC to the verification of respiratory motion. The method was parameterized to detect deviations from typical breathing motion. We analyzed the performance on synthetic data and on clinical episodes showing large correlation error. In comparison, we considered the prediction error of different state-of-the-art methods based on least mean squares (LMS; normalized LMS, nLMS; wavelet-based multiscale autoregression, wLMS), recursive least squares (RLSpred) and support vector regression (SVRpred).

Results

On synthetic data, OMC outperformed wLMS by at least 30 % and SVRpred by at least 141 %, detecting 70 % of transitions. No artifacts were detected by nLMS and RLSpred. On patient data, OMC detected 23–49 % of the episodes correctly, outperforming nLMS, wLMS, RLSpred and SVRpred by up to 544, 491, 408 and 258 %, respectively. On selected episodes, OMC detected up to 94 % of all events.

Conclusion

OMC is able to detect changes in breathing as well as artifacts which previously would have gone undetected, outperforming prediction error-based detection. Synthetic data analysis supports the assumption that prediction is very insensitive to specific changes in breathing. We suggest using OMC as an additional safety measure ensuring reliable and fast stopping of irradiation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Schweikard A, Glosser G, Bodduluri M, Murphy M, Adler J (2000) Robotic motion compensation for respiratory movement during radiosurgery. Comput Aided Surg 5(4):263–277

    Article  CAS  PubMed  Google Scholar 

  2. Schweikard A, Shiomi H, Adler J (2004) Respiration tracking in radiosurgery. Med Phys 31(10):2738–2741

    Article  PubMed  Google Scholar 

  3. Seppenwoolde Y, Berbeco RI, Nishioka S, Shirato H, Heijmen B (2007) Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: a simulation study. Med Phys 34(7):2774–2784

    Article  PubMed  Google Scholar 

  4. Depuydt T, Haas OC, Verellen D, Erbel S, De Ridder M, Storme G (2010) Geometric accuracy evaluation of the new VERO stereotactic body radiation therapy system. In: UKACC international conference on control 2010, pp 1–6, IET. doi:10.1049/ic.2010.0291

  5. Murphy MJ, Isaakson M, Jalden J (2002) Adaptive filtering to predict lung tumor motion during free breathing. In: CARS 2002 computer assisted radiology and surgery. Springer, Berlin, pp 539–544

  6. Ernst F, Schlaefer A, Schweikard A (2007) Prediction of respiratory motion with wavelet-based multiscale autoregression. Med Image Comput Comput Assist Interv—MICCAI 2007. In: 10th international conference on medical image computing and computer assisted intervention, vol 10, pp 668–675

  7. Ernst F, Schweikard A (2008) Prediction of respiratory motion using a modified recursive least squares algorithm. In: CURAC 2008-Tagungsband, pp 157–160

  8. Ernst F, Dürichen R, Schlaefer A, Schweikard A (2013) Evaluating and comparing algorithms for respiratory motion prediction. Phys Med Biol 58(11):3911–3929. doi:10.1088/0031-9155/58/11/3911

    Article  CAS  PubMed  Google Scholar 

  9. Ernst F, Schweikard A (2009) Forecasting respiratory motion with accurate online support vector regression (SVRpred). Int J Comput Assist Radiol Surg 4(5):439–447

    Article  PubMed  Google Scholar 

  10. Dürichen R, Wissel T, Ernst F, Schweikard A (2013) Respiratory motion compensation with relevance vector machines. In: Medical image computing and computer-assisted intervention–MICCAI 2013. Springer, Berlin, pp 108–115

  11. Murphy M, Pokhrel D (2009) Optimization of an adaptive neural network to predict breathing. Med Phys 36(1):40–47

    Article  PubMed  Google Scholar 

  12. Ramrath L, Schlaefer A, Ernst F, Dieterich S, Schweikard A (2007) Prediction of respiratory motion with a multi-frequency based extended kalman filter. In: Proceedings of the 21st international conference and exhibition on computer assisted radiology and surgery (CARS07), vol 21

  13. Dürichen R, Wissel T, Ernst F, Schlaefer A, Schweikard A (2014) Multivariate respiratory motion prediction. Phys Med Biol 59(20):6043

    Article  PubMed  Google Scholar 

  14. Schweikard A, Shiomi H, Adler J (2005) Respiration tracking in radiosurgery without fiducials. Int J Med Robot 1(2):19–27

    Article  CAS  PubMed  Google Scholar 

  15. Alur R, Dill D (1994) A theory of timed automata. Theor Comput Sci 126(2):183–235

    Article  Google Scholar 

  16. David A, Du D, Larsen KG, Legay A, Mikučionis M, Bøgsted Poulsen D, Sedwards S (2012) Statistical model checking for stochastic hybrid systems. In: Proceedings first international workshop on hybrid systems and biology, Newcastle Upon Tyne, 3rd September 2012, Electronic proceedings in theoretical computer science, vol 92. Open Publishing Association, pp 122–136

  17. Bengtsson J (2002) Clocks, DBMs and states in timed systems. Ph.D. thesis, Uppsala University

  18. Legay A, Delahaye B, Bensalem S (2010) Statistical model checking: an overview. In: Barringer H, Falcone Y, Finkbeiner B, Havelund K, Lee I, Pace G, Rou G, Sokolsky O, Tillmann N (eds) Runtime verification, vol 6418. Lecture Notes in Computer Science, Springer, Berlin, pp 122–135

  19. Behrmann G, David A, Larsen KG (2006) A tutorial on Uppaal 4.0. In: Bernardo M, Corradini F (eds) Formal methods for the design of real-time systems, vol 3185. Lecture Notes in Computer Science, Springer, Berlin, pp 1–48. Updated version of the original paper from 2004

  20. Bulychev P, David A, Larsen KG, Mikučionis M, Bøgsted Poulsen D, Legay A, Wang Z (2012) UPPAAL-SMC: statistical model checking for priced timed automata. In: Wiklicky H, Massink M (eds) 10th workshop on quantitative aspects of programming languages and systems (QAPL 2012), electronic proceedings in theoretical computer science, vol 85. Open Publishing Association, pp 1–16

  21. Bengtsson J, Yi W (2004) Timed automata: semantics, algorithms and tools. In: Desel J, Reisig W, Rozenberg G (eds) Lectures on concurrency and petri nets, vol 3098. Lecture Notes in Computer Science, Springer, Berlin, pp 87–124

  22. Li T, Tan F, Wang Q, Bu L, Cao JN, Liu X (2012) From offline toward real-time: a hybrid systems model checking and CPS Co-design approach for medical device plug-and-play (MDPnP). In: 2012 IEEE/ACM third international conference on cyber-physical systems. IEEE, pp 13–22

  23. Antoni S, Rinast J, Schupp S, Schlaefer A (2015) Comparing model-free motion prediction and on-line model checking for respiratory motion management. In: Gemeinsamer Tagungsband der Workshops der Tagung Software Engineering 2015, Dresden, Germany, 17.–18. März 2015, pp 15–18

  24. Antoni ST, Rinast J, Schupp S, Schlaefer A (2015) Evaluation des Einflusses von Artefakten auf den Korrelationsfehler in der bewegungskompensierten Radiochirurgie. In: CURAC 2015 - Tagungsband. Bremen, Germany, pp 133–138

  25. Krilavicius T, Wan K, Lee K, Man KL (2012) Specification and verification of radiation therapy system with respiratory compensation using uppaal. Int J Des Anal Tools Integr Circ Syst 3(1):8–13

    Google Scholar 

  26. Rzezovski N, Ernst F (2008) Graphical tool for the prediction of respiratory motion signals. In: CURAC 2008 - Tagungsband, pp 179–180

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Sven-Thomas Antoni.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Antoni, ST., Rinast, J., Ma, X. et al. Online model checking for monitoring surrogate-based respiratory motion tracking in radiation therapy. Int J CARS 11, 2085–2096 (2016). https://doi.org/10.1007/s11548-016-1423-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-016-1423-2

Keywords

Navigation